Measuring Fluoride in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Similar documents
HI Titratable Total Acidity Mini Titrator for Wine Analysis

HI Formol Number Mini Titrator for Wine and Fruit Juice Analysis

INSTRUCTION MANUAL Mi456

INSTRUCTION MANUAL Mi455

HI FREE & TOTAL SULPHUR DIOXIDE MINITITRATOR for wine analysis

LABORATORY INVESTIGATION

iworx Sample Lab Experiment GB-4: Ecological Balance

Acidity and ph Analysis

5KEK1322 W A_v08.indd 1 5/13/16 2:25 PM

Operating the Rancilio Silvia after PID kit modification Version 1.1

Calibrate grill heat zones Monthly GR 1 M1

INSTRUCTION MANUAL Mi453 Reducing Sugars

Lusso / Espresso Si' Models:PL-16 PAB-16. Made In Italy

Copyright JnF Specialties, LLC. All rights reserved worldwide.

User guide supplement. Built-in oven NZ AU. Soft Touch electronic models

WE CAN HELP. Smart Coffee Maker User Manual. Can t connect? Need help? DO NOT RETURN THIS PRODUCT TO THE STORE

INSTALLATION QUICK START GUIDE AND CHECKLIST ATTENTION!! DO NOT ATTEMPT A PARTIAL INSTALL.

OPERATING MANUAL. Sample PRO 100 Series. Electric Heating. Applies to Versions: SPE1*, SPE2, SPE4, SPE6

Beer bitterness and testing

+ = Power up your Smart Cup while pressing the corresponding button to reach different program modes. Heat Exchange fill/tank Drain Page:

Alcohol Meter for Wine. Alcolyzer Wine

MBAA DISTRICT ST. LOUIS. Home Brewing System. Rev 1.5 3/6/2010

LAB: One Tube Reaction Part 1

1. IMPORTANT SAFEGUARDS When using electrical appliances, basic safety precautions should always be followed to reduce the risk of fire, electric

INSTRUCTION MANUAL Milwaukee Wine Lab Photometer Mi454 Tartaric Acid

Installation & Parts Manual 2" & 3" High Capacity Strainers For Class 1, Petroleum Applications

Installation and User Instructions Flushing System Kit

phd Pitcher UltraWater Owner s Manual Delicious Great Tasting Antioxidant Alkaline Energized

CafeRomatica NICR7.. Fully automatic coffee centre Operating Instructions and Useful Tips. A passion for coffee.

Programming Guide For EXTRACTOR Coffee Brewers (CBS 2000e Series)

Find out all you need to know about measuring SO 2. in wine making

TRIPLE INFUSION PUMP WITH AUTOMATIC CYCLING TIMER

S m a rtset Control for Convection Models

I. INTRODUCTION I ITEMS:

MODEL: POD BREWER AURORA SAF 68 HOMEMAKER ESPRESSO MAKER WITH COFFEE POD

SERVICE MANUAL ESPRESSO COFFEE BREWER UNITS

Machine No. 2, SV=104. Figure 1

The fermentation of glucose can be described by the following equation: C6H12O6 2 CH3CH2OH + 2 CO2 + energy glucose ethanol carbon dioxide.

For Jura parts, repairs & tech support, contact: Services Unlimited, Inc./ Parts Guru

Given a realistic scenario depicting a new site install, the learner will be able to install and setup the brewer for retail turnover without error.

PROMOTIONS PROMOTIONS. A World of Expertise, in your Hands

Armonia Smart. Quick Start Guide MACHINE DESCRIPTION STARTING UP

CMC DUO. Standard version. Table of contens

LIFE COMES IN MANY FLAVORS

QUICK START GUIDE F10 - F20

Hand Crank Ice Shaver VKP1099

INSTALLATION & OPERATING GUIDE

BZ35 Espresso Machine

THE COBURN COMPANY, INC. PO Box 147 Whitewater, WI U.S.A. Phone: Fax:

A N F I M Srl. Caimano: Timer / Automatico / Manuale Super Caimano: Timer / Automatico Caimano SPecial 450: Timer / Automatico

Vinmetrica SC-50 MLF Analyzer User Manual

Vinmetrica SC-100A User Manual

YOUR MATERIAL HANDLING NEEDS

Density Measurement of Petroleum Products by Hydrometer Method ASTM D287 - ASTM D IP API ASTM D6822

Uno xl electronic. Turnspit.

Analytical Testing for Winemakers. Test crucial parameters in must and juice. hannainst.com

EBI 10 High Precision Wireless Data Loggers

Vinmetrica SC-100A User Manual

MTN Products. AquaCafé r (Touch Capsules) Installation & Operations Manual

INSTALLATION and OPERATION MANUAL for GXD SERIES BREWERS

Sleeve Down: Frothing. Sleeve Up: Steaming

Optimal solutions for your success.

Activity Instructions

... A Vital Part of Connection World CMPTER. Cable Connectors Receptacles With Solder End Protective Caps N CONNECTORS

Scishare Coffee Machine

Alcolyzer Plus Spirits

The Separation of a Mixture into Pure Substances

Wega Mininova 2003 Installation Instructions

A N F I M Srl. Caimano On Demand. Super Caimano COFFEE GRINDER C O D Y S C O D Y - II. Model

Magnetic Luminex Assay

Station 1. Polarity of Water

Copyright JnF Specialties, LLC. All rights reserved worldwide.

E N T E R P R I S E S

TOASTER OVEN USER MANUAL MODEL: PKMFT039

Lab 2-1: Measurement in Chemistry

HANDMADE IN FLORENCE GS/3. Software Programming Guide. Version 1.09

ICE CREAM MAKER DE LUXE INSTRUCTION MANUAL & RECIPE BOOK ICE-1611

CERTOMAT IS For controlled benchtop incubation

MODEL# GCM4500 COFFEE MAKER WITH GRINDER. PHOTO OF PRODUCT

The Column Oven Oven capabilities Oven safety Configuring the oven Making a temperature-programmed run Fast chromatography

Introduction to the General Chemistry II Laboratory. Lab Apparatus and Glassware

Westec 4-Channel. Micro-Oxygenation Unit. The Ox Box

Scanomat TopBrewer Daily Cleaning

K Mini Plus. Use & Care Guide. Get the most from your new Keurig brewer

Manual Spark Ignition (MSI) Kit For Use only with Thermocouple Manual Safety (TMSI) UL Approved Fire Pit Pans with Burning Spur

Table of Contents: The Basics Technical overview Char-Broil App...7. How to use your smoker Cooking modes...

Burr Grinder. Automatic burr grinder with 17-position grind selector. Instruction Booklet EM0430

Get Started. Better Everyday ECAM26455

ELECTRONIC OVEN CONTROL GUIDE

Determination of Alcohol Content of Wine by Distillation followed by Density Determination by Hydrometry

INSTALLATION MANUAL TG - ROTISSERIE OVEN MODELS

12 CUP COFFEE MAKER User Guide

» Red wine mash flooder FD-MÜ

Higher Resolution Separation of Organic Acids and Common Inorganic Anions in Wine

Espresso Multi Capsule Machine

IKAWA App V1 For USE WITH IKAWA COFFEE ROASTER. IKAWA Ltd. Unit 2 at 5 Durham Yard Bethnal Green London E2 6QF United Kingdom

Coffeemaker. Coffeemaker Safety... 2 Parts and Features... 3 Using Your Coffeemaker... 4 Cleaning Your Coffeemaker... 4 Customer Service...

RESOLUTION OIV-OENO ANALYSIS OF VOLATILE COMPOUNDS IN WINES BY GAS CHROMATOGRAPHY

MODELS 1315 & 1315-S La Pavoni CLUB COMBO

CAIMANO SPECIAL450 BARISTA

Transcription:

Measuring Fluoride in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 502 Key Words Thermo Scientific Orion Dual Star ph/ise meter, fluoride ion selective electrode (ISE), fluoride, wastewater, water Goal This technical note explains how to accurately measure fluoride content in general water and wastewater samples using the Thermo Scientific Orion Dual Star ph/ise dual channel benchtop meter. Introduction Too much or too little fluoride can be harmful to humans and the environment. Precise measurement of fluoride by municipalities, potable water plants and bottled water manufacturers promotes quality water for consumption and helps prevent damage to the environment. Required Equipment Orion Dual Star ph/ise meter with electrode stand (Cat. No. 2115000) Fluoride combination electrode (Cat. No. 9609BNWP) Automatic Temperature Compensation () probe (Cat. No. 927007MD) Stirrer probe (Cat. No. 096019) or magnetic stir plate and bar 1 ppm fluoride standard premixed with Total Ionic Strength Adjustment Buffer (TISAB) II (Cat. No. 040906) 2 ppm fluoride standard premixed with (Cat. No. 040907) solution (Cat. No. 940909) Thermo Scientific Orion Optimum Results A electrode filling solution (Cat. No. 900061) Deionized water 100 ml volumetric flasks 50 ml beakers 50 ml, 25 ml and 15 ml pipettes 50 ml graduated cylinders or 10 ml automatic pipette

2 Fluoride Electrode Setup Note: Add filling solution each day before using the electrode. The filling solution level should be at least one inch above the level of sample to ensure a proper flow rate. The fill hole should always be open when taking measurements. See Figure 1 for labeled electrode components. Calibration Standard Preparation 1. Prepare a 0.5 ppm fluoride standard by pipetting 50 ml of the 1 ppm fluoride standard premixed with (Cat. No. 040906) and 25 ml of the TISAB II solution (Cat. No. 940909) into a 100 ml volumetric flask. Dilute to the mark with deionized water. Mix the solution well. 1. Remove the protective shipping cap from the sensing element and save the cap for storage. 2. Insert the spout of the Optimum Results A bottle into the electrode fill hole, add a small amount of filling solution to the chamber, tip the electrode upsidedown to wet the O-ring and return the electrode to its original, vertical position. 3. Hold the electrode body and use your thumb to push down on the electrode cap to allow a few drops of filling solution to drain out and wet the junction. 1 ppm F - with 50 ml 50 ml of 1 ppm 100 ml volumetric flask H 2 0 25 ml 25 ml of 4. Release the electrode cap and allow the outer sleeve to return to its original position. If the outer sleeve does not return to its original position immediately, check if the O-ring is moist and repeat step 3. 0.5 ppm 5. Add filling solution to the electrode up to the bottom of the electrode fill hole. 0.5 ppm 6. Soak the electrode in the 1 ppm fluoride calibration standard until all of the equipment is ready for use. 2. Using a graduated cylinder or automatic pipette, add of the 0.5 ppm fluoride standard to a 50 ml beaker and label the beaker. Cap Spring Cable 0.5 ppm 0.5 ppm Outer body Filling chamber O-ring Fill hole 3. Using a graduated cylinder or automatic pipette, add each of the 1 ppm fluoride standard and 2 ppm fluoride standard to separate 50 ml beakers and label the beakers. Inner body Reference element Inner cone 1 ppm F - with 1 ppm 2 ppm F - with 2 ppm Figure 1 : Fluoride Electrode Components

Sample Preparation 1. Using a graduated cylinder or pipette, measure and add 15 ml of the sample to a 50 ml beaker and label the beaker. Add 15 ml of to the beaker. Gently stir the sample and to mix the solution. 15 ml of sample 15 ml of Meter Setup Note: It is highly recommended that the EZ Startup menu be completed the first time that the meter is used. The EZ Startup menu sets important meter parameters, such as the displayed language, date and time, measurement mode and read type for each channel, and data output settings. To access the EZ Startup menu from the measurement mode, press the setup key, press the / keys to highlight EZ Startup and press the f2 (select) key. 1. In the measurement mode, press the setup key. 3 Sample 1 Sample 1 or 2. Press the / keys to highlight Channel 1 or Channel 2, depending on which BNC input the fluoride electrode was connected to, and press the f2 (select) key. 2. Repeat step 1 for additional samples. Meter Preparation The Orion Dual Star meter has two BNC, reference and connections. These connections are labeled as Channel 1 or Channel 2 on the ridge above the connections. Channel 2 Channel 1 3. Press the / keys to highlight Measure Mode and highlight ISE and press the f2 (accept) key. Press the / keys to highlight ppm (or mg/l, depending on your preference or method) and press the f2 (accept) key. 4. Press the / keys to highlight Electrode ID and highlight F - and press the f2 (accept) key. BNC Ref. USB Ref. Stirrer Ground Power BNC RS232 5. Press the / keys to highlight Resolution and press the f2 (select) key. Press the / keys to highlight 3 Significant Figures (1.00) and press the f2 (accept) key. 1. Prepare the power adapter by selecting the appropriate wall outlet plug and sliding the plug plate into the groove on the back of the adapter. 2. Connect the power adapter to the meter and then to the wall outlet. See Figure 2. Connect the fluoride electrode to one of the BNC inputs on the meter and note which channel (channel 1 or channel 2) was selected. See Figure 3. Connect the probe to the 8 pin MiniN input on the meter and note which channel was selected. See Figure 4. Connect the stirrer probe to the stir jack input on the meter. See Figure 5. und Ref. Power. wer BNC RS232 Ground Re Stirrer Figure 2 Figure 3 Figure 4 Figure 5 Gro 6. Press the / keys to highlight Temperature Input and highlight 1 or 2, depending on which 8 pin MiniN input the probe was connected to, and press the f2 (accept) key. 7. Press the / keys to highlight Calibration Setup and highlight Autoblank and press the f2 (select) key. Press the / keys to highlight On and press the f2 (accept) key. Verify that the Low Level Stability and Isopotential options are set to Off. Press the f1 (back) key. 8. Press the / keys to highlight Read Type and press the f2 (select) key. Press the / keys to highlight On Ready and press the f2 (select) key. 9. Press the f1 (back) key. 10. Press the / keys to highlight Instrument Parameters and press the f2 (select) key. 11. Press the / keys to highlight Export Data (PC/ Printer/Log) and press the f2 (select) key. Press the / keys to highlight Export Trigger and press the f2 (select) key. Press the / keys to highlight Channel 1 or Channel 2, depending on which BNC input the fluoride electrode was connected to, and press the f2 (accept) key. Press the / keys to highlight Data Log and press the f2 (select) key. Press the / keys to highlight Data Log On/Off and highlight On and press the f2 (accept) key.

4 12. Press the f1 (back) key twice. 13. Press the / keys to highlight Stirrer Speed and highlight 1 and press the f2 (accept) key. Press the f1 (back) key. 7. Press the f2 (next) key to proceed to the next calibration standard. 8. Rinse the fluoride electrode, probe and stirrer probe with deionized water, blot dry and place into the 1 ppm calibration standard. 14. Press the f1 (back) key. The meter will return to the measurement mode. Calibration Procedure 1. Prepare and condition the fluoride electrode. Connect the fluoride electrode, probe and stirrer probe to the meter (note which channel the fluoride electrode is connected to) and place the electrode and probes in the electrode stand. Prepare the 2 ppm, 1 ppm and 0.5 ppm standards. 1 ppm 9. When the electrode and standard are ready, press the f3 (start) key. 2 ppm 0.5 ppm 2. In the measurement mode, press the cal key. 3. Dual channel display only: Press the / keys to highlight Channel 1 or Channel 2 as the channel to calibrate and press the f2 (accept) key. select... Channel 1 Channel 2 Calibratio 1 ppm 4. Rinse the fluoride electrode, probe and stirrer probe with deionized water, blot dry and place into the 0.5 ppm calibration standard. 10. Wait for the concentration value to stop flashing and then use the numeric keypad and the decimal key to enter the concentration of the second standard as 1.00 and press the f2 (accept) key. Ch.1 ppm F _ Accept, or enter standard value P1 0.50 25.0 C Std. 2 = 1.00ppm 11. Press the f2 (next) key to proceed to the next calibration standard. 12. Rinse the electrode, probe and stirrer probe with deionized water, blot dry and place into the 2 ppm calibration standard. 0.5 ppm 2 ppm 5. When the electrode and standard are ready, press the f3 (start) key to begin the calibration. 13. When the electrode and standard are ready, press the f3 (start) key. 6. Wait for the concentration value to stop flashing and then use the numeric keypad and the decimal key to enter the concentration of the first standard as 0.50 and press the f2 (accept) key. Ch.1 ppm F _ Accept, or enter standard value Std. 1 = 0.50 ppm 14. Wait for the concentration value to stop flashing and then use the numeric keypad and the decimal key to enter the concentration of the third standard as 2.00 and press the f2 (accept) key. F Accept, or enter standard value P1 0.50 25.0 C P2 1.00 25.0 C Std. 3 = 2.00ppm

15. Press the f3 (cal done) key. A summary of the calibration will be displayed. The slope should be -54 mv per decade to -65 mv per decade. 5. Repeat steps 2 through 4 for all of the samples. 5 Sample x F - 16. Press the f2 (log/print) key to save and end the calibration and export the calibration data to the calibration log. NO3-6. When all of the samples have been measured, store the equipment. For short term storage between samples and overnight, store the fluoride electrode in the 1 ppm fluoride calibration standard. Store the probe and stirrer probe dry. Sample Measurement Procedure 1. Calibrate the fluoride electrode and meter (note which channel the fluoride electrode is connected to) and make sure that the electrode and probes are in the electrode stand. Prepare the fluoride samples. Sample 1 Sample 2 Sample 3 Sample 4 2. Rinse the fluoride electrode, probe and stirrer probe with deionized water, blot dry and place into the sample. Press the stirrer key to turn on the stirrer probe. Sample 1 3. The meter display will flash stabilizing and then show ready once the measurement is stable. Record the concentration and temperature of the sample when the meter display shows ready. When ready is shown on the display, the meter will export the measurement to the data log. Dual Channel 1 NH3 NH 3 or 1 Single Channel Display Mode / / 4. Press the stirrer key to turn off the stirrer probe. Remove the fluoride electrode, probe and stirrer probe from the sample.

Visit www.thermoscientific.com/water to view the full line of Thermo Scientific Orion products and download product literature, user guides and manuals and software updates. For any questions or technical support, contact your local Technical Sales Representative or our Technical Support Specialists by phone at 1-978-232-6000 or 1-800-225-1480 (US toll-free), or by email at wai.techservbev@thermofisher.com. Ordering Information To purchase an Orion Dual Star meter, electrodes and other related products, please contact your local equipment distributor and reference the part numbers listed below. Product Description Part Number Meter Orion Dual Star ph/ise Dual Channel Benchtop Meter 2115000 Electrodes Solutions Accessories Orion Fluoride Combination Electrode Orion Stainless Steel Temperature Probe 9609BNWP 927007MD Orion 1 ppm Fluoride Standard Premixed with, 475 ml 040906 Orion 2 ppm Fluoride Standard Premixed with, 475 ml 040907 Orion 10 ppm Fluoride Standard Premixed with, 475 ml 040908 Orion Solution, 3.8 L (1 gallon) 940909 Orion Optimum Results A Electrode Filling Solution, 5 x 60 ml 900061 Orion Stirrer Probe 096019 Swing Arm Electrode Stand 090043 Technical Note 502 thermoscientific.com/water 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. Water and Lab Products North America Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 info.water@thermofisher.com TN-FLWATER-E 0615 RevB Germany Tel: (49) 6184-90-6000 info.water.uk@thermofisher.com China Tel: (86) 21-68654588 India Tel: (91) 22-4157-8800 Singapore Tel: (65) 6778-6876 Japan Tel: (81) 045-453-9175 Australia Tel: (613) 9757-4300 In Australia: (1300) 735-295 InfoWaterAU@thermofisher.com