EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

Similar documents
HOW MUCH DYE IS IN DRINK?

Introduction to the General Chemistry II Laboratory. Lab Apparatus and Glassware

Royal Society of Chemistry Analytical Division East Anglia Region National Schools' Analyst Competition

Copyright JnF Specialties, LLC. All rights reserved worldwide.

Introduction to Measurement and Error Analysis: Measuring the Density of a Solution

Application Note FP High Sensitivity Coumarin Analysis. Introduction. Keywords

Practical 1 - Determination of Quinine in Tonic Water

Experiment 6 Thin-Layer Chromatography (TLC)

Gravimetric Analysis

Gravimetric Analysis

Lab 2-1: Measurement in Chemistry

Chemistry 212 MOLAR MASS OF A VOLATILE LIQUID USING THE IDEAL GAS LAW

SYNTHESIS OF SALICYLIC ACID

PECTINASE Product Code: P129

Separating the Components of a Mixture

Setting up your fermentation

I. INTRODUCTION I ITEMS:

3. Aspirin Analysis. Prelaboratory Assignment. 3.1 Introduction

Caffeine in Energy Drinks

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF UNBRUSHED 325 MESH GRIT, SLURRY METHOD: T

Experiment 2: ANALYSIS FOR PERCENT WATER IN POPCORN

Analytical Method for Coumaphos (Targeted to agricultural, animal and fishery products)

Determination of Alcohol Content of Wine by Distillation followed by Density Determination by Hydrometry

CHEM Experiment 4 Introduction to Separation Techniques I. Objectives

Paper Chromatography and Steam Distillation of Orange Oil EVERY STUDENT MUST BRING AN ORANGE TO LAB FOR THIS EXPERIMENT! Equipment

Correlation of the free amino nitrogen and nitrogen by O-phthaldialdehyde methods in the assay of beer

Rapid Analysis of Soft Drinks Using the ACQUITY UPLC H-Class System with the Waters Beverage Analysis Kit

Separating the Components of a Mixture

Chapter 14 Tex-619-J, Analysis of Water for Chloride and Sulfate Ions

Experiment 3: Separation of a Mixture Pre-lab Exercise

Mastering Measurements

Separation of a Mixture

IPPA SAG method revision 0,

Lab 2. Drug Abuse. Solubility and Colligative Properties of Solutions: Coffee, Soda, and Ice Cream

Determination of the concentration of caffeine, theobromine, and gallic acid in commercial tea samples

Lab 2. Drug Abuse. Solubility and Colligative Properties of Solutions: Coffee, Soda, and Ice Cream

LAB: One Tube Reaction Part 1

HI-1017: Pharmacy Technician. Module 10 Pharmacy Equipment

EXTRACTION PROCEDURE

ASSET EZ4-NCO Dry Sampler Extraction Procedure.

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

Analysis of tea powder for adulterant

TISSUE PREPARATION FOR LASER MICRODISSECTION (LCM) WITH ETHANOL:ACETIC ACID FIXATION

2. Other constituents in the sample solution should not interfere with the precipitation of the component of interest.

15. Extraction: Isolation of Caffeine from Tea

EXTRACTION OF SEDIMENTS FOR BUTYLTINS

LABORATORY INVESTIGATION

Average Matrix Relative Sensitivity Factors (AMRSFs) for X-ray Photoelectron Spectroscopy (XPS)

Determination of Metals in Wort and Beer Samples using the Agilent 5110 ICP-OES

Measuring tannins in grapes and red wine using the MCP (methyl cellulose precipitable tannin assay

Specific Heat of a Metal

Experimental Procedure

Elemental Analysis of Yixing Tea Pots by Laser Excited Atomic. Fluorescence of Desorbed Plumes (PLEAF) Bruno Y. Cai * and N.H. Cheung Dec.

Mathur Agar This medium is made up of the following reagents: dextrose, magnesium sulfate, potassium phosphate, neopeptone, yeast extract, and agar.

Separations. Objective. Background. Date Lab Time Name

Student Handout Procedure

Volume NaOH ph ph/ Vol (ml)

Analysis of Beta-Carotene and Total Carotenoids from Pacific Sea Plasma (Spectrophotometric Method)

Lab 2: Phase transitions & ice cream

Roast Beef Jan. 1, 2009

Heron Bay Ultra Premium 6 Week Wine Kit

An Economic And Simple Purification Procedure For The Large-Scale Production Of Ovotransferrin From Egg White

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

Title: Reagents Copy No: ## SOP No.: 3.13/3.1/S Effective Date: July 8, 2013 Location: ### Reagents

THIN LAYER CHROMATOGRAPHY AND MELTING POINT DETERMINATION: DETECTION OF CAFFEINE IN VARIOUS SAMPLES

EXTRACTION. Extraction is a very common laboratory procedure used when isolating or purifying a product.

QUANTITATIVE ASSAY FOR OCHRATOXIN A IN COFFEE, COCOA, AND SPICES (96-well kit)

INSTRUCTION MANUAL Mi453 Reducing Sugars

WATER AND SOIL PROPERTIES. ExPERImENTAL PROCEdURE

Bromine Containing Fumigants Determined as Total Inorganic Bromide

The Separation of a Mixture into Pure Substances

Unit Test: Nature of Science

Separating the Components of a Mixture

The grade 5 English science unit, Solutions, meets the academic content standards set in the Korean curriculum, which state students should:

1. Blender: Osterizer, 10-speed, or equivalent. 2. Separatory Funnel: Kilborn or equivalent (see figure 1) 2. HCl Solution: HCl/water (7:93 by volume)

Application Note: Analysis of Melamine in Milk (updated: 04/17/09) Product: DPX-CX (1 ml or 5 ml) Page 1 of 5 INTRODUCTION

Application Note No. 184/2015

Density Gradient Column Lab

Coffee-and-Cream Science Jim Nelson

Vinmetrica s SC-50 MLF Analyzer: a Comparison of Methods for Measuring Malic Acid in Wines.

Measure the specific heat of lead. Identify an unknown metal from its specific heat (optional),

Conceptual Chemistry. Conceptual Chemistry Glow In The Forensic Investigations

National Food Safety Standard

Acidity and ph Analysis

! " # # $% 004/2009. SpeedExtractor E-916

Teacher Notes on Explore: Can You Find the Floaters and Sinkers?

Certified Home Brewer Program. Minimum Certification Requirements

FED. TEST METHOD STD. 191A AND SHALL BE RETAINED UNTIL SUCH CHANGE NOTICE 3 TIME AS THE STANDARD IS REVISED. January 15, 1986

Experimental Procedure

C27 Chromatography. Collect: Column Mortar and pestle Dropper (229 mm) Capillary tube TLC plate Aluminum foil UV light

COMPARATIVE EVALUATION OF CLARIFYING REAGENTS OCTAPOL AND LEAD SUB ACETATE FOR USE WITH MASSECUITES AND MOLASSES. Niconor Reece and Sydney Roman

A COMPARATIVE STUDY OF THE CAFFEINE PROFILE OF MATURE TEA LEAVES AND PROCESSED TEA MARKETED IN SONITPUR DISTRICT OF ASSAM, INDIA.

A Simple Guide to TDS Testing

Figure 11.1 Derivatives of Salicylic Acid O C OH OCH3. Na + OH sodium salicylate. OH CH3 Acetylsaliclic acid Aspirin.

Validation Report: Total Sulfite Assay Kit (cat. no. K-TSULPH)

Figure 1 Fluorescence Fingerprint of Pineapple Juice and Prediction of Autofluorescence Substances

EXPERIMENT NO. 3 HYDROMETER ANALYSIS ASTM D-422

Ochratoxin A N H. N-{ [(3R)-5-chloro-8-hydroxy-3-methyl-1-oxo-3,4-dihydro-1H-isochromen-7-yl]carbon yl}- L-phenylalanine

EXTRACTION OF SEDIMENTS FOR AROMATIC AND CHLORINATED HYDROCARBONS

Standard Operating Procedure (SOP) for Quartz dissolution in Hydrofluoric Acid

Transcription:

EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay UNKNOWN Submit a clean, labeled 500-mL volumetric flask to the instructor so that your unknown quinine solution may be issued. Your name, section number, and your locker number should be written legibly on this flask. The flask does not need to be dry on the inside, but needs to have been rinsed with distilled water after it has been cleaned. The flask must be turned in at least 1 lab period before you plan to do the experiment so that the Teaching Assistants will have time to prepare the unknown. Each student will have his or her own unknown to analyze even if you are working in pairs. BACKGROUND Quinine (C 20 H 24 N 2 O 2, 324.43 g/mol) is an alkaloid extracted from the bark of the cinchona tree. It has been used for many years as an antimalarial agent. Although it does not cure malaria, it is effective in alleviating the symptoms of malarial attacks. The usual medicinal form is quinine dihydrochloride or quinine sulfate dihydrate, (C 20 H 24 N 2 O 2 ) 2 H 2SO 4 2H 2O, 782.97 g/mol. Quinine is a very strongly fluorescing compound, especially in dilute acid solution, and thus can be detected in very trace amounts. In 0.05 M H 2 SO 4, quinine has two analytically useful excitation wavelengths: λ ex = 250 and 350 nm. Regardless of which excitation wavelength is used, the wavelength of maximum fluorescence emission intensity, λ em or λ fl, is 450 nm. The basis for quantitation is that the intensity of fluorescence emission in very dilute solutions is directly proportional to the concentration of quinine if the intensity of the excitation source and other experimental factors are kept constant. Because the absolute emission intensity can vary considerably with small differences in experimental conditions, a calibration curve is prepared by measuring the fluorescence-emission intensity of accurately known quinine standard solutions. INSTRUMENTATION Turner Quantech Digital Filter Fluorometer, Model FM 109525 The instrument used in this experiment uses glass filters to select wavelength ranges appropriate for quinine, rather than some type of expensive grating or prism monochromator to isolate narrow excitation and fluorescence wavelengths. The latter type may typically have a bandpass of 0.1 or 1 nm. A bandpass is defined as the width at half-height of the maximum intensity of light that passes through the filter. It is also called the bandwidth. CHE 226 Analytical Chemistry Laboratory 35 Quinine Fluorescence

A monochromator-based instrument is much better for obtaining spectra, particularly those with fine structure, but a filter-based unit will often be better for routine quantitative analysis. The much larger bandpasses permit greater fractions of the excitation and fluorescent light to excite the sample and to reach the detector, respectively. This greatly increases the sensitivity of the instrument and can thus usually lower typical detection by an order of magnitude or more. Filter fluorometers are also much less expensive. Excitation Source: A 5-watt quartz-halogen lamp, which emits intense broadband radiation from 340 nm to 750 nm. Excitation Wavelength Filter: A narrow-band 360-nm filter with a bandpass of 40 nm. Emission Wavelength Filter: A sharp cut-in, long-wavelength-pass filter which transmits essentially all light with λ > 415 nm. Detector: Photomultiplier tube, model 931B PMT. Detection Limit. Stated in the manufacturer s literature as 30 ppt quinine sulfate, which is 30 ng/l or about 9 x 10-11 M. Turn the instrument on at least 15 minutes before using to allow it to warm up and stabilize. The ON-OFF switch is on the back panel near the power cord. When turned on, the instrument runs a countdown timer during which it undergoes self tests. PREPARATION OF SOLUTIONS Preparation of Stock Sulfuric Acid Solutions The two stock H 2 SO 4 solutions should have already been prepared for you by the Teaching Assistants. However, if you use up all of the 0.05 M solution while doing your experiment, prepare another 2-L batch for those that follow. It is very straightforward. 1 M H 2 SO 4. Slowly and carefully add 56 ml conc. H 2 SO 4 to about 500 ml deionized water in a 1-L beaker with stirring. This solution should have been already prepared for your use. 0.05 M H 2 SO 4. With a graduated cylinder, add 100 ml of 1 M H 2 SO 4 with a graduated cylinder to about 500 ml of distilled water in the screw-capped acid reagent jug labeled for this solution. Mix and dilute to the 2.0-L mark on the bottle and mix thoroughly. This solution should already have been prepared for your use. If you use all of or most of what is there, prepare a new batch for the students who follow you. Preparation of Quinine Stock Solution, 1000 ppm 1. Carefully weigh exactly 0.1207 g of quinine sulfate dihydrate onto a folded glassine weighing paper or into a small plastic weighing boat, and transfer this quantitatively into a 100-mL volumetric flask. A few squirts of distilled water from a wash bottle should help to wash the solid material from the weighing boat and the neck of the flask. CHE 226 Analytical Chemistry Laboratory 36 Quinine Fluorescence

2. Pipet 5.00 ml of 1 M H 2 SO 4 (located in hood #2) into the flask. Carefully dissolve all the quinine in this sulfuric acid solution by swirling before diluting to volume. This is critically important. 3. Carefully dilute to volume with distilled water and mix thoroughly. Preparation of Intermediate Quinine Stock Solution, 10.0 ppm 1. Pipet 5.00 ml of the 1000-ppm solution into a 500-mL volumetric flask. 2. Add 25.0 ml of 1 M H 2 SO 4, dilute carefully to volume with distilled water, and mix thoroughly. Preparation of Quinine Standard Solutions 1. Using volumetric transfer pipets and/or a 10-mL graduated pipet, add 1.00, 3.00, 5.00, 7.50, and 10.00 ml of the 10-ppm intermediate stock solution into five properly labeled 100-mL volumetric flasks. This will result in standard solutions of 0.1, 0.3, 0.5, 0.75, and 1.0 ppm. 2. Carefully dilute to volume with 0.05 M H 2 SO 4. 3. 0.05 M H 2 SO 4 is used as the blank. Preparation of Quinine Unknown 1. Your unknown solution is obtained from the teaching assistants in a 500-mL volumetric flask. Add 25.0 ml of 1 M H 2 SO 4 to the flask. 2. Dilute to volume with distilled water and mix thoroughly. PROCEDURE Turn on the Quantec fluorometer at least 15-20 minutes before making measurements in order to let it warm up and stabilize. Measurement of Emission Intensities 1. Carefully fill separate, clean, plastic fluorescence cuvettes (these have 4 clear sides) about ¾ full with the blank (0.05 M H 2 SO 4 ), each of the five standards, and each unknown sample. Do not touch the optical surfaces with your fingers. Instead, handle with KimWipes. Be sure to wipe any smudges off the optical surfaces of the cuvettes. 2. On the main menu of the fluorometer, press ENTER. CHE 226 Analytical Chemistry Laboratory 37 Quinine Fluorescence

3. Use the right cursor ( ) to select Quinine. Press RETURN. 4. The instrument asks if filters are correct. Press RETURN. 5. The instrument asks for the number of points for the calibration curve. Use the up-arrow ( ) to increase the number to 5 in order to be able to use your five standards. Press RETURN. 6. Beginning with the highest concentration standard, enter the concentration using the uparrow to change the values. Use ( ) to move the cursor to the appropriate position. Use the up-arrow to set the appropriate units (ppm). 7. The instrument asks to insert the sample. Do so and press RETURN. 8. Repeat for other standards. 9. Instrument asks the operator to insert the BLANK. Do so and press RETURN. The instrument displays READ and gives the value. 10. Insert the unknown and Press RETURN. 11. Readout shows READ and provides the value. 12. Repeat the entire process at least two more times to obtain a set of three good replicate measurements: If the values for a set of standards and the unknown do not drift over time, this would mean that the raw fluorescence data for your unknown are within a range of about 3-4% relative. If the values do appear to drift, then the ratios of the fluorescence intensities of the standard to those of a nearby standard are within a range of about 3-4% relative. Comparison of Calibration Procedures This procedure will compare values that are (a) calculated by the instrument based on its internal calibration procedure with (b) those obtained based on the calibration curve(s) that you prepare from the raw fluorescence-emission data obtained above. 1. At main menu, select RAW FLUORESCENCE as the units of measure. 2. The instrument requests a value for the standard. Enter the value and RETURN. 3. The instrument instructs you to insert the standard. USE the 1.0 ppm STANDARD. Do so and press RETURN. The instrument adjusts its internal gain. 4. The instrument asks Insert Blank Sample? Select YES, followed by RETURN. 5. Insert the blank and the instrument will take a reading. CHE 226 Analytical Chemistry Laboratory 38 Quinine Fluorescence

6. Now insert the standard solutions and the unknown, and record the emission in fluorescence units for each. 7. Repeat the entire procedure at least twice more in order to obtain at least 3 good measurements for each sample. 8. Prepare calibration curves and calculate the concentration of the unknown sample. When done with the entire experiment, rinse out all the plastic cuvettes with distilled water, shake off excess water, and return them to the experiment drawer so they can be reused. DATA ANALYSIS Prepare appropriate calibration curve(s) from your replicate sets of data. Depending on the nature of the data and any drift in the instrument, it may be best to (a) average all the net emission intensities for a particular solution and to prepare one calibration curve or (b) prepare several calibration curves from the separate sets of data, obtain several values for the concentration of quinine in your unknown, then average these values. Determine which one is better or if both give equivalent results. Use linear least squares to fit each of the three sets of data, and the average of all three, in your analysis. Report the best value for the concentration of quinine (in ppm) in your unknown and the estimated standard deviation of this best value. Note: By definition, parts-per-million (ppm) is a mass-ratio unit g analyte /g sample x 10 6 or µg/g. In dilute aqueous solutions, 1 ppm is equivalent to 1 µg/ml or 1 mg/l because the density of the solution is 1.00 g/ml. A 1000-ppm solution of quinine sulfate (1.000 g of quinine sulfate per liter of solution) would only be an 828.7 ppm of quinine itself. If you cannot weigh out exactly 0.1207 g of quinine sulfate, try to get it close, and simply calculate the correct values of the concentrations of your standards. HAZARDOUS WASTE DISPOSAL Dispose of ALL waste quinine and sulfuric acid solutions in the proper, labeled Hazardous Waste Container for this experiment, which is located in the hoods. 0.05 M H 2 SO 4 has a ph of about 1 and thus may NOT simply be run down the drain, even if you run large amounts of water with it. If you are unsure about the container, ASK. TEXT REFERENCE D. A. Skoog, D. M. West, F. J. Holler, and S. R. Crouch, Analytical Chemistry: An Introduction, 7th ed., Chapter 23, pp. 594-631. Revised January 24, 2005 Copyright by The Department of Chemistry,, 2005 CHE 226 Analytical Chemistry Laboratory 39 Quinine Fluorescence