AppNote 13/2012. Automated Solid Phase Extraction (SPE)-LC-MS/MS Method for the Determination of Acrylamide in Brewed Coffee Samples KEYWORDS ABSTRACT

Similar documents
AppNote 4/2003. Fast Analysis of Beverages using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

Extraction of Acrylamide from Coffee Using ISOLUTE. SLE+ Prior to LC-MS/MS Analysis

AppNote 2/2003. Wine Discrimination using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 1/2003. Fast Analysis of Food and Beverage Products using a Mass Spectrometry Based Chemical Sensor KEYWORDS ABSTRACT

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

AppNote 13/2002. Classifi cation of Coffees from Different Origins by Chemical Sensor Technology INTRODUCTION

Solid Phase Micro Extraction of Flavor Compounds in Beer

Application Note: Analysis of Melamine in Milk (updated: 04/17/09) Product: DPX-CX (1 ml or 5 ml) Page 1 of 5 INTRODUCTION

Extraction of Multiple Mycotoxins From Animal Feed Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

AppNote 6/2013. Using TDU-Pyrolysis-GC-MS to Investigate Aged Whiskey Samples and Their Oak Barrels ABSTRACT INTRODUCTION

Solid Phase Micro Extraction of Flavor Compounds in Beer

Determination of Caffeine in Coffee Products According to DIN 20481

Rapid Analysis of Soft Drinks Using the ACQUITY UPLC H-Class System with the Waters Beverage Analysis Kit

Determination of Pesticides in Coffee with QuEChERS Extraction and Silica Gel SPE Cleanup

The Determination of Pesticides in Wine

High Sensitivity Quantitation Method of Dicyandiamide and Melamine in Milk Powders by Liquid Chromatography Tandem Mass Spectrometry

Automation of AOAC for the Identification of FD&C Color Additives in Foods using Solid Phase Extraction

Vinmetrica s SC-50 MLF Analyzer: a Comparison of Methods for Measuring Malic Acid in Wines.

! " # # $% 004/2009. SpeedExtractor E-916

Tyler Trent, SVOC Application Specialist; Teledyne Tekmar P a g e 1

Determination of Methylcafestol in Roasted Coffee Products According to DIN 10779

Fast Analysis of Smoke Taint Compounds in Wine with an Agilent J&W DB-HeavyWax GC Column

Zoe Grosser, Vinson Leung, Jim Fenster, Brian LaBrecque Horizon Technology, Inc., Salem, NH USA

Determination of Metals in Wort and Beer Samples using the Agilent 5110 ICP-OES

Determination of natamycin in wines Résolution OIV-SCMA

Determination of the concentration of caffeine, theobromine, and gallic acid in commercial tea samples

Determination of Ochratoxin A in Roasted Coffee According to DIN EN 14132

Rapid Tea Analysis on Poroshell 120 SB-C18 with LC/MS

High-Resolution Sampling 2D-LC with the Agilent 1290 Infinity II 2D-LC Solution

Higher Resolution Separation of Organic Acids and Common Inorganic Anions in Wine

ASSET EZ4-NCO Dry Sampler Extraction Procedure.

Speciated Arsenic Analysis in Wine Using HPLC-ICP-QQQ

EXTRACTION PROCEDURE

EXTRACTION OF SEDIMENTS FOR AROMATIC AND CHLORINATED HYDROCARBONS

Analytical Method for Coumaphos (Targeted to agricultural, animal and fishery products)

Application Note CL0311. Introduction

Enhancing the Flexibility of the NGC Chromatography System: Addition of a Refractive Index Detector for Wine Sample Analysis

Analysis of trace elements and major components in wine with the Thermo Scientific icap 7400 ICP-OES

RAPID, SPECIFIC ANALYSIS OF MELAMINE CONTAMINATION IN INFANT FORMULA AND LIQUID MILK BY UPLC/MS/MS

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

Somchai Rice 1, Jacek A. Koziel 1, Anne Fennell 2 1

Improved Cleanup of Pesticides in Dry, Difficult Matrices Using a Novel Dual-Layer SPE Cartridge for LC/MS/MS and GC/MS/MS Analysis

APPLICATIONS TN Fast and Robust Analysis of Organic Acids from Wine using HPLC-UV. Introduction. Results and Discussion. Materials and Methods

Correlation of the free amino nitrogen and nitrogen by O-phthaldialdehyde methods in the assay of beer

EXTRACTION OF SEDIMENTS FOR BUTYLTINS

General overview of the two stages of the QuEChERS technique. Stage 1: Sample extraction. Stage 2: Sample cleanup

Caffeine in Energy Drinks

Determination of Pesticide Residues in Red Wine

RESOLUTION OIV-OENO ANALYSIS OF VOLATILE COMPOUNDS IN WINES BY GAS CHROMATOGRAPHY

TSKgel TECHNICAL INFORMATION SHEET No. 131

Use of Accelerated Solvent Extraction (ASE) with Centrifugal Evaporation to Automate Fat Determination in Chocolate

CHAPTER 8. Sample Laboratory Experiments

Somchai Rice 1, Jacek A. Koziel 1, Jennie Savits 2,3, Murlidhar Dharmadhikari 2,3 1 Agricultural and Biosystems Engineering, Iowa State University

three different household steam ovens, representing a number of identically constructed ovens (see attached list at the end of this document):

A COMPARATIVE STUDY OF THE CAFFEINE PROFILE OF MATURE TEA LEAVES AND PROCESSED TEA MARKETED IN SONITPUR DISTRICT OF ASSAM, INDIA.

Profiling of Aroma Components in Wine Using a Novel Hybrid GC/MS/MS System

Table 1: Experimental conditions for the instrument acquisition method

Fast Analysis of Arsenic Species in Wines using LC-ICP-QQQ

Extraction of Phenolic Acids from Plant Tissue Using Accelerated Solvent Extraction (ASE )

PECTINASE Product Code: P129

Metabolomic Profiling of Wines using LC/QTOF MS and MassHunter Data Mining and Statistical Tools

Identification of Adulteration or origins of whisky and alcohol with the Electronic Nose

Analytical Report. Volatile Organic Compounds Profile by GC-MS in Cupcake Batter Flavor Concentrate

Bromine Containing Fumigants Determined as Total Inorganic Bromide

Alcohol Meter for Wine. Alcolyzer Wine

QUANTITATIVE ASSAY FOR OCHRATOXIN A IN COFFEE, COCOA, AND SPICES (96-well kit)

[ application note note ] ]

Identification and Semiquantitation of Monoterpene Glycosides in Ripening Muscat of Alexandria Grapes

Analytical Report. Volatile Organic Compounds Profile by GC-MS in Clove E-liquid Flavor Concentrate. PO Box 2624 Woodinville, WA 98072

A Fast Method for Sugar Analysis of Instant Coffee Samples

WORK ORDER NUMBER:

One class classification based authentication of peanut oils by fatty

AppNote 4/2000. Flavor Profi ling of Beverages by Stir Bar Sorptive Extraction (SBSE) and Thermal Desorption GC/MS/PFPD KEYWORDS ABSTRACT

Ochratoxin A N H. N-{ [(3R)-5-chloro-8-hydroxy-3-methyl-1-oxo-3,4-dihydro-1H-isochromen-7-yl]carbon yl}- L-phenylalanine

Determination Of Saponin And Various Chemical Compounds In Camellia Sinensis And Genus Ilex.

Separations. Objective. Background. Date Lab Time Name

Protect Your Mass Spec

ANALYTICAL SUMMARY REPORT

Unique Carbohydrate Profiles In Different Brands of Tequila

Profiling and Identification of Hop-Derived Bitter Compounds in Beer Using LC-MS/MS

AWRI Refrigeration Demand Calculator

Analysis of Resveratrol in Wine by HPLC

Application Note FP High Sensitivity Coumarin Analysis. Introduction. Keywords

Samples: Standard solutions of rutin, quercetina, rosmarinic acid, caffeic acid and gallic acid. Commercial teas: Green, Ceilan, Hornimans and Black.

Call for AOAC INTERNATIONAL Collaborator

powder and cocoa butter and various spices in the range of 1-20 ppb (µg/kg).

Hongwei Xie, Martin Gilar, Asish Chakraborty, Weibin Chen, and Scott Berger Waters Corporation, Milford, MA, U.S. EXPERIMENTAL

Assessment of the CDR BeerLab Touch Analyser. March Report for: QuadraChem Laboratories Ltd. Campden BRI Group contracting company:

Abstract. Introduction

Supplementary Material

EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

SUPPLEMENTARY MATERIALS. Methylxanthine content in commonly consumed foods in Spain and determination of its intake during consumption

Practical 1 - Determination of Quinine in Tonic Water

Estimation of Caffeine Concentration in Decaffeinated Coffee and Tea Available in Pakistan

The espresso ritual. Product Overview: Semi-automatic coffee machines.

Identification & Estimation of Melamine Residue in Powdered Milk by RP-HPLC

Fingerprinting Analysis of Different Types of Beer Using Comprehensive 2D-LC

Analytical Report. Table 1: Target compound levels. Concentration units are ppm or N/D, not detected.

ClearFlux. Dialyzer Regeneration System

Transcription:

AppNote 13/2012 Automated Solid Phase Extraction (SPE)-LC-MS/MS Method for the Determination of Acrylamide in Brewed Coffee Samples Fredrick D. Foster, John R. Stuff, and Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA KEYWORDS Solid Phase Extraction, LC-MS/MS, Sample Analysis, Lab Automation, Coffee ABSTRACT Acrylamide is thought to be produced during the roasting process associated with coffee production. Acrylamide has been labeled as a probable human carcinogen. Due to the use of roasted coffee beans in making coffee and the high consumption of coffee world-wide, brewed coffee could be a source of daily exposure to acrylamide. Acrylamide determination has been shown to be challenging due to presence of coextractives in the final extract. Manual solid phase extraction followed by LC-MS/MS analysis has been reported as a successful method for the determination of acrylamide from brewed coffee samples. However, performing solid phase extraction manually can be tedious and time consuming and there is increasing demand for automation of these methods. In this study, we show that a manual SPE method used for the determination of acrylamide in brewed coffee can be converted to an autosampler-compatible cartridge format and automated using a robotic autosampler controlled by user-friendly software. Calibration standards prepared in freshly brewed green coffee (unroasted) resulted in a linear calibration curve (r 2 =0.99) from 1 ng/ml to 500 ng/ml. Precision of the automated SPE-LC/MS/MS method was calculated as CV = 1.7 %.

INTRODUCTION Solid phase extraction (SPE) is one of the most popular sample preparation methods employed by chromatographers, based on the numerous published SPE methods found in the literature. Typically, a liquid sample is passed across an adsorbent bed to retain and concentrate target analytes and eliminate interference from the sample matrix. Alternatively, the adsorbent can be used to retain interferences while allowing the target analytes to pass through. Previous work has shown that since standard SPE cartridge sorbents are used with the GERSTEL SPE Option, established manual SPE procedures can be transferred directly, automated, and conveniently controlled by the MAESTRO software [1]. A solid phase extraction procedure for the separation of acrylamide from brewed coffee has previously been shown to be effective for subsequent accurate determination by LC-MS/MS [2]. This SPE method takes advantage of both the ability of an adsorbent bed to retain and concentrate the target analytes and eliminate interference from the sample matrix as well as the ability of a second adsorbent bed to be used to retain interferences while allowing the target analytes to pass through. For illustrative purposes, Figure 1 shows a graphical representation of a typical automated SPE extraction process. After optimization of the SPE sorbent selection, the manual SPE procedure was translated into an automated SPE method. EXPERIMENTAL Materials. Acrylamide (A3553-100G) was purchased from Sigma-Aldrich. D3-acrylamide (DLM-821-0.1), used as internal standard, was purchased from Cambridge Isotope Laboratories, Inc. Independent 1 mg/ml stock solutions of acrylamide and D3- acrylamide were prepared using 0.1 % formic acid in water. Intermediate acrylamide stock solutions were prepared by appropriately diluting the analyte stock solution with 0.1 % formic acid in water. A working internal standard stock solution was prepared at a concentration of 10 μg/ml using 0.1 % formic acid in water. All ground coffee samples were brewed using an electric drip coffee maker. Ground coffee was added to a coffee filter (4.8 g/cup) and cold tap water was added to the water reservoir. After brewing, 10 ml of brewed coffee was transferred to a 10 ml glass screw top vial to be used for analysis. Matrix matched calibration standards were prepared by making appropriate dilutions of the intermediate acrylamide stock solutions using freshly brewed green coffee (confirmed to have no acrylamide initially) to give concentrations from 1 ng/ml to 500 ng/ml. Following preparation, 10 ml of each standard was transferred to a 10 ml glass screw top vial to be used for analysis along with the other brewed coffee samples. 3 ml Strata-X SPE cartridges and 3 ml AccuCAT SPE cartridges, both in the GERSTEL SPE format, were placed in series onto the appropriate autosampler tray. Instrumentation. All automated SPE PrepSequences were performed using a MultiPurpose Sampler (MPS XL) in the Dual Head configuration fitted with the GERSTEL SPE Option as shown in Figure 2. All analyses were performed using an Agilent 1290 HPLC with a Phenomenex Synergi Hydro-RP column (2.0 x 250 mm, 4 μm, 80 Å), an Agilent 6460 Triple Quadrupole Mass Spectrometer with Jet stream electrospray source and GERSTEL MPS XL autosampler configured with an Active Wash Station. Sample injections were made using a 6 port (0.25 mm) Cheminert C2V injection valve outfitted with a 10 μl stainless steel sample loop. Figure 1. Graphical representation of a typical automated SPE procedure. AN/2012/13-2

RESULTS AND DISCUSSION The manual steps in the published SPE procedure [2] include: Figure 2. MultiPurpose Sampler (MPS XL) with GERSTEL SPE Option. Analysis conditions LC. Pump: flowrate = 0.2 ml/min Mobile Phase: A - 5 mm formic acid in water B acetonitrile Gradient: isocratic, 5 % B Run time: 10 min Injection volume: 10 μl (loop over-fill technique) Column temp.: 45 C Analysis conditions MS. Operation: electrospray positive mode Gas temp.: 350 C Gas flow (N 2 ): 5 L/min Nebulizer pressure: 35 psi Sheath Gas Heater: 250 C Sheath Gas Flow: 11 L/min Capillary voltage: 4000 V Delta EMV: +700 V MS/MS parameters. D3 Acrylamide: m/z 75.1 -> 58.1 Acrylamide: m/z 72.1 -> 55.1 Dwell: 100 msec Fragmentor voltage: 52 V Collison energy: 8 V Cell accelerator voltage: 7 V 1. A 3 ml/200 mg Phenomenex Strata-X SPE cartridge is conditioned using 3.5 ml of methanol followed by 3.5 ml of water. 2. 1.5 ml of brewed coffee sample is loaded onto the conditioned SPE cartridge and allowed to pass through the sorbent material, followed by 0.5 ml of water. 3. The cartridge is eluted with 1.5 ml of water and the eluate collected for further cleanup. 4. A 3 ml/200 mg AccuCAT SPE cartridge is conditioned with 2.5 ml of methanol followed by 2.5 ml of water. 5. 0.5 ml of the eluate collected from Strata-X cartridge is loaded onto the conditioned SPE cartridge and allowed to pass through the sorbent material to waste. 6. 0.8 ml of the eluate collected from the Strata-X cartridge is loaded onto the SPE cartridge and allowed to pass through the sorbent material and the eluate is collected. 7. 10 μl of the collected eluate from step 6. is injected for LC-MS/MS analysis. We first determined that the electric drip coffee maker was free of acrylamide by taking the appropriate amount of cold tap water through a brewing process without the presence of ground coffee. The resulting water was then processed using the SPE procedure and analyzed using the LC-MS/MS method. No acrylamide was detected from the extracted water sample. The manual SPE procedure was translated into a MAESTRO Prep Sequence and was demonstrated to successfully prepare sample extracts for subsequent analysis by LC-MS/MS. The Phenomenex Strata- X sorbent was used as a recommended alternative SPE sorbent since, during the course of method development, it was found that samples processed using the Strata-X sorbent resulted in higher peak area responses when compared to samples processed using the original SPE sorbent [2]. AN/2012/13-3

Figure 3 shows, (from left to right), 1) a brewed coffee sample, 2) the Phenomenex Strata-X, 3 ml/200 mg, SPE cartridge in the GERSTEL format after completing step 3 of the automated SPE procedure, 3) the eluate resulting from step 3 of the automated SPE procedure, 4) the Varian Bond Elut AccuCAT, 3 ml/200 mg, SPE cartridge in the GERSTEL format after completing step 6 of the automated SPE procedure, and 5) the eluate resulting from step 6 of the automated SPE procedure. As shown in Figure 3, the automated SPE method provides a clear eluate for LC-MS/MS analysis. 1 2 Figure 3. Representative picture showing an example brewed coffee sample at the various steps of the SPE procedure. 3 4 5 Figure 4 shows an overlay of the mass chromatograms for all calibration curve samples. The limit of quantitation was found to be 1 ng/ml acrylamide in brewed green coffee. Figure 4. Overlay of mass chromatograms from calibration curve samples. AN/2012/13-4

The calibration curve is shown in Figure 5. As can be seen, calibration standards prepared in freshly brewed green coffee (unroasted) resulted in a linear calibration curve (r 2 =0.99) from 1 ng/ml to 500 ng/ml. Figure 5. Acrylamide calibration curve. Accuracy and precision of the automated SPE method was assessed using six independently prepared 25 ng/ml acrylamide QC samples. As shown in Table 1, the average % accuracy was found to be 97.9 % and the precision of the automated SPE-LC- MS/MS method was calculated as CV = 1.7 %. Table 1. Automated SPE-LC-MS/MS: accuracy and precision results (Peak Area Ratio) Calc. % Accuracy Replicate 1 0.264 99.3 Replicate 2 0.267 100 Replicate 3 0.260 97.8 Replicate 4 0.255 96.1 Replicate 5 0.256 96.5 Replicate 6 0.259 97.7 mean 0.260 97.9 SD 0.00433 1.63 % CV 1.66 1.67 Figure 6 shows a comparison of a blank water sample, an extracted blank brewed green coffee sample, and an extracted 1 ng/ml brewed green coffee calibration standard. No acrylamide was detected for an extracted blank brewed green coffee sample. A B C Figure 6. Comparison of blank water sample (A), extracted blank brewed green coffee sample (B), and extracted 1 ng/ml brewed green coffee calibration standard (C). AN/2012/13-5

Figure 7 shows a comparison of an extracted 1 ng/ml acrylamide calibration standard in brewed green coffee to representative mass chromatograms from two (2) extracted blank brewed coffee samples (Brand A and Brand B). Acrylamide was observed to be present in both of these coffee samples. All samples were analyzed with and without the addition of internal standard. A B C Figure 7. Comparison of extracted 1 ng/ml acrylamide calibration standard in brewed green coffee (A), extracted blank brewed brand A coffee sample (B), and extracted blank brewed brand B coffee sample (C). To ensure that the SPE-LC-MS/MS method would result in accurate determination of the concentration of acrylamide in brewed coffee samples, independent of matrix differences, a standard was added to a blank brewed coffee sample, spiking it with 10 ng/ml acrylamide and then analyzing it in triplicate. The blank brewed sample was also analyzed in triplicate. As shown in Table 2, the resulting difference between the calculated acrylamide concentration from the extracted blank samples and that of the extracted spiked samples was found to be 11 ng/ml, indicating average recovery of 110 %. Table 2. Results of acrylamide spiking experiment. Acrylamide Sample Name Response AN/2012/13-6 D3-Acrylamide (ISTD) Response Calc. Conc. Blank Subtr. Calc. Conc. Brewed coffee (Run 1) 9966 211092 4.84 - Brewed coffee (Run 2) 8686 201278 4.40 - Brewed coffee (Run 3) 7904 191375 4.19 - Brewed coffee spiked with 10 ng/ml (Run 1) 28465 201476 15.1 10.7 Brewed coffee spiked with 10 ng/ml (Run 2) 29514 207499 15.2 10.8 Brewed coffee spiked with 10 ng/ml (Run 3) 31121 207502 16.1 11.6 mean 11.01 SD 0.522 % CV 4.74 % Recovery 110

A total of eleven (11) different commercial coffee samples were brewed and then analyzed in duplicate using the automated SPE-LC-MS/MS method. As shown in Table 3, acrylamide was detected in all but one of the brewed coffee samples examined. Table 3. Brewed coffee analyses results. Brewed Coffee Sample Average Calc. Conc. [ng/ml] Brand A 4.21 Brand B 3.45 Brand C 3.14 Brand D 2.17 Brand E - Brand F 2.76 Brand G 1.97 Brand H 5.63 Brand I 6.21 Brand J 2.42 Brand K 5.84 CONCLUSIONS As a result of this study, we were able to show: A manual SPE procedure could be readily automated using the Dual Head GERSTEL MPS XL sampler and SPE sorbents in cartridges formatted for automation. A linear calibration curve resulting in an r 2 value of greater than 0.99 was achieved with a limit of quantitation for acrylamide of 1 ng/ml. The SPE-LC-MS/MS method proved to be accurate and precise. Accuracy data averaged 97.9 % (range: 96.1 % - 100 %) and a precision result of CV = 1.67 % for acrylamide in QC samples was achieved. Acrylamide was found in ten of the eleven brewed coffee samples analyzed using the automated SPE- LC/MS/MS method. REFERENCES [1] F. Foster, J. Stuff, J. Whitecavage, and E. Pfannkoch, Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control, AppNote 3/2009, www.gerstelus.com [2] D. Andrzejewski, J. Roach, M. Gay, and S. Musser, Analysis of Coffee for the Presence of Acrylamide by LC-MS/MS J. Agric. Food Chem. 2004, 52, 1996-2002. AN/2012/13-7

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability