Determination of Metals in Wort and Beer Samples using the Agilent 5110 ICP-OES

Similar documents
Analysis of trace elements and major components in wine with the Thermo Scientific icap 7400 ICP-OES

Application note. Determination of metals in wine using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer. Food Testing and Agriculture

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

Uses of profiling trace metals in wine with ICP- MS and Mass Profiler Professional (MPP) for the wine industry

The Impact of Vineyard Origin and Winery on the Elemental Profile of Red Wines

Application Note FP High Sensitivity Coumarin Analysis. Introduction. Keywords

Determination of Caffeine in Coffee Products According to DIN 20481

*Level IV report narratives are more detailed than other levels.

Solid Phase Micro Extraction of Flavor Compounds in Beer

Rapid Analysis of Soft Drinks Using the ACQUITY UPLC H-Class System with the Waters Beverage Analysis Kit

Solid Phase Micro Extraction of Flavor Compounds in Beer

Harvest Series 2017: Wine Analysis. Jasha Karasek. Winemaking Specialist Enartis USA

Application Note: Analysis of Melamine in Milk (updated: 04/17/09) Product: DPX-CX (1 ml or 5 ml) Page 1 of 5 INTRODUCTION

Speciated Arsenic Analysis in Wine Using HPLC-ICP-QQQ

High Sensitivity Quantitation Method of Dicyandiamide and Melamine in Milk Powders by Liquid Chromatography Tandem Mass Spectrometry

Elemental Analysis of Yixing Tea Pots by Laser Excited Atomic. Fluorescence of Desorbed Plumes (PLEAF) Bruno Y. Cai * and N.H. Cheung Dec.

Determination of Methylcafestol in Roasted Coffee Products According to DIN 10779

Determination of Ochratoxin A in Roasted Coffee According to DIN EN 14132

IMPROVING THE PROCEDURE FOR NUTRIENT SAMPLING IN STONE FRUIT TREES

Extraction of Acrylamide from Coffee Using ISOLUTE. SLE+ Prior to LC-MS/MS Analysis

Correlation of the free amino nitrogen and nitrogen by O-phthaldialdehyde methods in the assay of beer

EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

ALPHA. Innovation with Integrity. FT-IR Wine & Must Analyzer FT-IR

RESOLUTION OIV-OENO ANALYSIS OF VOLATILE COMPOUNDS IN WINES BY GAS CHROMATOGRAPHY

Vinmetrica s SC-50 MLF Analyzer: a Comparison of Methods for Measuring Malic Acid in Wines.

Application Note CL0311. Introduction

Analytical Method for Coumaphos (Targeted to agricultural, animal and fishery products)

DETERMINATION OF CHROMIUM IN MACEDONIAN WINE BY ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY

Alcohol Meter for Wine. Alcolyzer Wine

Fast Analysis of Arsenic Species in Wines using LC-ICP-QQQ

Tyler Trent, SVOC Application Specialist; Teledyne Tekmar P a g e 1

Allergens in wine a specific detection of Casein, Egg and Lysozyme

PECTINASE Product Code: P129

OenoFoss. Instant quality control throughout the winemaking process. Dedicated Analytical Solutions

AppNote 13/2012. Automated Solid Phase Extraction (SPE)-LC-MS/MS Method for the Determination of Acrylamide in Brewed Coffee Samples KEYWORDS ABSTRACT

Keywords Green and black tea. Infusions. Sample preparation. Multi-element analysis. Principal component analysis. Linear discriminant analysis

Determination of natamycin in wines Résolution OIV-SCMA

Appendix B: Monitoring methods, accuracy, detection limits and precision (updated for 2003)

! " # # $% 004/2009. SpeedExtractor E-916

Determination of the concentration of caffeine, theobromine, and gallic acid in commercial tea samples

Operating the Rancilio Silvia after PID kit modification Version 1.1

Chemometric analysis of minerals and trace elements in Sicilian wines from two

Enhancing the Flexibility of the NGC Chromatography System: Addition of a Refractive Index Detector for Wine Sample Analysis

The Determination of Pesticides in Wine

RIDASCREEN Gliadin. Validation Report. R-Biopharm AG. Art.No. R7001

Exploring Attenuation. Greg Doss Wyeast Laboratories Inc. NHC 2012

Measuring Fluoride in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Production, Optimization and Characterization of Wine from Pineapple (Ananas comosus Linn.)

Emerging Applications

Automation of AOAC for the Identification of FD&C Color Additives in Foods using Solid Phase Extraction

CHAPTER 8. Sample Laboratory Experiments

Assessment of the CDR BeerLab Touch Analyser. March Report for: QuadraChem Laboratories Ltd. Campden BRI Group contracting company:

WORK ORDER NUMBER:

High-Resolution Sampling 2D-LC with the Agilent 1290 Infinity II 2D-LC Solution

Extraction of Multiple Mycotoxins From Animal Feed Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Fast Analysis of Smoke Taint Compounds in Wine with an Agilent J&W DB-HeavyWax GC Column

CONCENTRATION LEVELS OF METALS IN COMMERCIALLY AVAILABLE ETHIOPIAN ROASTED COFFEE POWDERS AND THEIR INFUSIONS

Glutomatic System. Measure Gluten Quantity and Quality. Gluten Index: AACC/No ICC/No. 155&158 Wet Gluten Content: ICC/No.

An Investigation of Methylsufonylmethane as a Fermentation Aid. Eryn Bottens, Jeb Z Hollabaugh, and Thomas H. Shellhammer.

Determination of Pesticides in Coffee with QuEChERS Extraction and Silica Gel SPE Cleanup

Determination of Lead in Saudi Arabian Imported Green Tea by ICP-MS

Validation Report: Total Sulfite Assay Kit (cat. no. K-TSULPH)

Measuring tannins in grapes and red wine using the MCP (methyl cellulose precipitable tannin assay

VINEYARD NUTRIENTS AT BROOKWOOD ESTATE MARGARET RIVER, WESTERN AUSTRALIA

HI Formol Number Mini Titrator for Wine and Fruit Juice Analysis

ANALYTICAL SUMMARY REPORT

Acidity and ph Analysis

Recent Developments in Rheological Instruments

Gluten Index. Application & Method. Measure Gluten Quantity and Quality

TSKgel TECHNICAL INFORMATION SHEET No. 131

EXTRACTION OF SEDIMENTS FOR AROMATIC AND CHLORINATED HYDROCARBONS

Laboratory Performance Assessment. Report. Analysis of Pesticides and Anthraquinone. in Black Tea

Michigan Grape & Wine Industry Council Annual Report 2012

HOW MUCH DYE IS IN DRINK?

WineScan All-in-one wine analysis including free and total SO2. Dedicated Analytical Solutions

Alcolyzer Plus Spirits

three different household steam ovens, representing a number of identically constructed ovens (see attached list at the end of this document):

DRAFT TANZANIA STANDARD

EXTRACTION OF SEDIMENTS FOR BUTYLTINS

806 West Beacon Rd. Client Project #: PCSB-LEAD First Draw Lakeland,FL Date Sampled: Oct 3, 2018 Oct 9, 2018; Invoice:

Detecting Melamine Adulteration in Milk Powder

DETERMINATION OF CADMIUM IN SAUDI ARABIAN IMPORTED GREEN TEA SAMPLES BY ICP-MS

Identification of Adulteration or origins of whisky and alcohol with the Electronic Nose

Asian Journal of Chemistry; Vol. 23, No. 7 (2011),

OenoFoss Instant Quality Control made easy

Royal Society of Chemistry Analytical Division East Anglia Region National Schools' Analyst Competition

Analytical Report. Volatile Organic Compounds Profile by GC-MS in Clove E-liquid Flavor Concentrate. PO Box 2624 Woodinville, WA 98072

Average Matrix Relative Sensitivity Factors (AMRSFs) for X-ray Photoelectron Spectroscopy (XPS)

Higher Resolution Separation of Organic Acids and Common Inorganic Anions in Wine

Validation Report: Free Sulfite Assay Kit (cat. no. K-FSULPH)

Improved Cleanup of Pesticides in Dry, Difficult Matrices Using a Novel Dual-Layer SPE Cartridge for LC/MS/MS and GC/MS/MS Analysis

RAPID, SPECIFIC ANALYSIS OF MELAMINE CONTAMINATION IN INFANT FORMULA AND LIQUID MILK BY UPLC/MS/MS

Profiling of Aroma Components in Wine Using a Novel Hybrid GC/MS/MS System

Allergen analysis of food and surfaces with sensitive test kits

Application Note No. 184/2015

Certificate of Analysis

Analytical Report. Volatile Organic Compounds Profile by GC-MS in Cupcake Batter Flavor Concentrate

Fact Sheet ANALYSIS. Measuring total anthocyanins (colour) in red grape berries. Scope. Equipment and apparatus

Quality of western Canadian peas 2017

Somchai Rice 1, Jacek A. Koziel 1, Anne Fennell 2 1

Transcription:

Determination of Metals in Wort and Beer Samples using the Agilent 5110 ICP-OES Authors Application Food and Beverages Dana Sedin 1, Stacey Williams 1, Elizabeth Kulikov 2, Jenny Nelson 3, Greg Gilleland 3 1. New Belgium Brewing Company, Fort Collins, CO, USA 2. Agilent Technologies, Melbourne, Australia 3. Agilent Technologies, Santa Clara, CA, USA Introduction The analysis of metals during the beer brewing process is important as certain elements can affect the quality and taste of final beer products. Metals can originate from a range of sources including the brewing water, malt grains, hops, adjuncts, fruits, and spices. They can also be introduced through the brewing and packaging process. Brewers deliberately introduce metals in the form of salts (CaSO 4, MgSO 4, ZnSO 4, CaCl 2 ) to control ph, adjust taste, improve efficiency, and enhance fermentation performance. Metals that can be detrimental to the overall taste of beer include iron, which can impart a metallic taste. As a result, the concentration of Fe must be kept as low as possible in the finished product. Typical techniques used for the identification of metals in beers include Flame Atomic Absorption Spectroscopy (FAAS) and Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES).

The Agilent 5110 Vertical Dual View (VDV) ICP-OES is a suitable option for food and beverage testing laboratories. The instrument s vertically oriented torch and solid-state RF (SSRF) system, which operates at 27 MHz, deliver a stable and robust plasma that can handle a range of complex sample types. The 5110 ICP-OES uses a high-speed Vista Chip II CCD detector that provides good precision and fast analysis times. The high productivity of the 5110 ensures that argon consumption per sample is low. The instrument s large dynamic range of 8 orders allows the measurement of essential nutrients at high concentrations and minor elements at trace levels. This range of concentration levels is typical in food and beverage testing. In addition to quantification of selected elements, IntelliQuant, a semiquantitative feature in the ICP Expert software, can identify and calculate an approximate concentration for up to 70 elements in a sample. IntelliQuant allows analysts in beverage labs to monitor the additional elements in their samples, without changing their routine methods. IntelliQuant can acquire data across all wavelengths from 167 to 785 nm with a further reading of Agilent s VistaChip II CCD detector, in as little as 15 additional seconds per sample. This study focused on the determination of Ca, K, Mg, Na, Cu, Fe, Mn, and Zn in wort and finished beer samples using the Agilent 5110 VDV ICP-OES. Experimental Instrumentation All measurements were performed using an Agilent 5110 VDV ICP-OES. The sample introduction system consisted of a SeaSpray nebulizer, double-pass cyclonic spray chamber, and a 1.8 mm i.d injector torch. Using a combined run, some elements were measured in radial mode and others in axial. The user can set up multiple conditions in the ICP Expert software and for each sample analyzed, two sets of conditions are measured and the results are combined in a single report. Instrument operating parameters are shown in Table 1. Table 1. Agilent 5110 ICP-OES instrument and method parameters. Parameter Setting Cu Fe Mn Zn Ca K Mg Na Read time (s) 10 5 Replicates 3 Sample uptake delay (s) 35 Stabilization time (s) 25 3 Rinse time (s) 35 Pump speed (rpm) 13 Fast pump during uptake and rinse (rpm) On RF power (kw) 1.20 Auxillary flow (L/min) 1.00 Plasma flow (L/min) 12.0 Nebulizer flow (L/min) 0.70 Viewing mode* Axial Radial Viewing height (mm) - 8 Sample pump tubing Internal standard pump tubing Waste pump tubing Background correction *Combined run of two viewing modes. Black/black Orange/green Blue/blue Fitted Standards and sample preparation Wort and beer samples were obtained from the New Belgium Brewing Company (Fort Collins, CO, USA). Sample details are listed in Table 2. All in-process and finished beer samples were sonicated for 20 mins to remove CO 2. Following sonication, 10 ml of beer was diluted in 10 ml of 14% HNO 3. All wort samples were filtered through Whatman paper before preparation. 8 ml of wort was diluted in 32 ml of 8.8% HNO 3 and 3.8% ethanol. All beer and wort samples were stored at 4 C before analysis. Table 2. New Belgium Brewing Company wort and finished beer samples. Sample code W1 W2 W3 W4 B1 B2 B3 B4 Sample description Voodoo Ranger IPA wort Fat Tire wort Voodoo Ranger 8 Hop Pale Ale wort Voodoo Ranger Imperial wort Voodoo Ranger IPA finished beer Fat Tire finished beer Voodoo Ranger 8 Hop Pale Ale finished beer Voodoo Ranger Imperial finished beer 2

Multi-elemental calibration standards were prepared at 0.1 and 0.5 ppm for Cu, Fe, Mn, and Zn; and at 15.0 and 50.0 ppm for Ca, K, Mg, and Na. All standards were prepared in 7% HNO 3 and 3% ethanol. Continuing Calibration Verification (CCV) standards were prepared at 0.02 ppm for minor element checks, 15 ppm for majors, and 500 ppm for K check. A control beer sample spiked with 0.1 ppm Zn was also prepared. Internal standards (ISTD), gallium and yttrium, were prepared at 200 ppm (Ga) and 40 ppm (Y). All calibration standards, Quality Control (QC) checks, and internal standards were matrix matched with 7% HNO 3 + 3% ethanol. Results and Discussion Detection limits The Detection Limit (DL) for each element was based on three sigma of seven replicate measurements of the blank solution during an analytical run (Table 3). Sub-ppm level DLs were achieved for all elements. Method DLs (MDLs) were determined by calculating 3 x the SD of 10 replicate beer and wort samples, respectively. The MDLs are also given in Table 3. Table 3. Element wavelengths, DLs, and MDLs. Element & wavelength DL Major elements Wort MDL Beer MDL Ca 422.673 0.0126 13.35 10.41 K 769.897 0.6539 57.35 66.93 Mg 279.553 0.0091 8.12 15.28 Na 589.592 0.0301 2.07 2.16 Minor elements Cu 327.395 0.0005 0.027 0.014 Fe 238.204 0.0008 0.061 0.018 Mn 259.372 0.0001 0.015 0.020 Zn 213.857 0.0009 0.034 0.016 Calibration linearity Linear calibrations were obtained for all elements, with calibration coefficients greater than 0.999 (Table 4) and less than 10% calibration error for each point. Figure 1 shows linear calibration curves for Fe, Cu, Mg, and Ca. Figure 1. Calibration curves for Fe 238.204 nm, Cu 327.395 nm, Mg 279.553 nm, and Ca 422.673 nm. 3

Table 4. Wavelength and working calibration range. Element and wavelength Ca 422.673 Background correction/ interference correction (Y 371.029) Calibration range Correlation coefficient 0 50 0.9999 K 769.897 (Ga) 0 50 0.9995 Mg 279.553 (Ga) 0 50 0.9996 Na 589.592 (Ga) 0 50 0.9999 Cu 327.395 Fe 238.204 (Y 360.074) (Y 371.029) 0 0.5 0.9999 0 0.5 1.0000 Mn 259.372 (Ga) 0 0.5 0.9999 Zn 213.857 Ga 417.204 Y 371.029, 360.074 (Y 360.074) Used as ISTD Used as ISTD 0 0.5 1.0000 Detectability test To validate the method, a series of QC spike recovery tests were carried out during each analytical run. The tests consisted of a Continuing Calibration Blank (CCB), followed by two CCV solutions (low concentration for Cu, Fe, Mn and Zn, high concentration for Ca, Mg and Na), a K check, and control beer sample spiked with 0.1 ppm Zn. Each QC solution was analyzed six times and averaged, with all recoveries within ±10% of the target values. The QC results are given in Table 5. Analysis of wort and finished beer samples The four wort and four finished beer samples were analyzed using the developed method. The quantified concentration results are displayed in Table 6. All the results for all elements were within the specification limits set by the manufacturer. Comparing the concentration levels of the elements in wort and beer enables the analyst to monitor the beer brewing process. The data is also useful to characterize the product. Table 5. QC spike recovery results of the low and high CCVs, and control samples. Ca 422.673 Cu 327.395 Fe 238.204 Element and wavelength K 769.897 Mg 279.553 Mn 259.372 Na 589.592 Zn 213.857 Low conc check 0.02 ppm (mean, n=6), CCV - 0.0215 0.0210 - - 0.0208-0.0205 % Recovery - 107 105 - - 104-103 Zinc check 0.1 ppm, control sample (mean, n=6) - - - - - - - 0.098 % Recovery - - - - - - - 98 High conc check 15 ppm (mean, n=6), CCV 14.758 - - 15.761 15.494-14.610 - % Recovery 98 - - 105 103-97 Potassium check 500 ppm (mean, n=6) - - - 464.014 - - - - % Recovery - - - 92 - - - - Table 6. Concentration of major and minor elements in wort and finished beer samples. Sample Ca 422.673 Cu 327.395 Fe 238.204 K 769.897 Mg 279.553 Concentration Mn 259.372 Na 589.592 Zn 213.857 W1 96 0.242 0.045 857 129 0.183 30 0.163 W2 80 0.195 0.149 675 111 0.150 26 0.167 W3 85 0.204 0.054 735 124 0.241 26 0.183 W4 74 0.357 0.055 1125 178 0.263 36 0.262 B1 90 0.099 0.029 719 134 0.287 25 0.016 B2 61 0.047 0.038 464 90 0.132 17 0 B3 101 0.076 0.023 692 106 0.308 12 0 B4 62 0.148 0.032 850 145 0.295 26 0.028 IntelliQuant semiquantitative results A wort and finished beer sample were analyzed using IntelliQuant during the analytical run. A custom IntelliQuant calibration (1.0, 5.0 and 10.0 mg/l) was created using an Agilent QC standard solution. The semiquantitative results for all elements in the wort and finished beer samples are shown in Table 7. The results reveal the presence of high levels of silicon in the samples. Silicon is present in barley and is introduced at high levels during the beer brewing process. Levels are typically not monitored during the process but can be of interest in terms of dietary intake. 4

Table 7. IntelliQuant semiquantitative results of elements in wort and finished beer samples. Element and wavelength Wort IntelliQuant value Finished beer IntelliQuant value Na 9.3 17.5 Mg 24.9 48.3 Si 3.35 17.6 K 136.1 309.5 Ca 20.4 30.6 Mn 0.1 0.1 Sr - 0.1 B 0.1 - Conclusions Major and minor metals in wort and finished beer samples were measured with good speed and accuracy using the Agilent 5110 VDV ICP-OES. With its vertically oriented torch and 27 MHz solid-state RF system, the robust 5110 is suited to the routine analysis of complex samples, such as alcoholic beverages. The method can be used to provide valuable information to brewers at different stages of production allowing them to optimize the quality of the final product. In addition to the quantified method results, high levels of Si were identified in the wort and beer samples using the IntelliQuant (semiquantitative) feature of the ICP Expert software. Acknowledgements Agilent Technologies would like to thank the New Belgium Brewing Company for their co-operation and collaboration. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2017 Published September 28, 2017 Publication number: 5991-8394EN