ETAAS DETERMINATION OF SOME TRACE ELEMENTS IN WINE

Similar documents
DETERMINATION OF CHROMIUM IN MACEDONIAN WINE BY ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY

Determination of cadmium in wine by electrothermal atomic absorption spectrometry


Application note. Determination of metals in wine using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer. Food Testing and Agriculture

Analysis of trace elements and major components in wine with the Thermo Scientific icap 7400 ICP-OES

Appendix B: Monitoring methods, accuracy, detection limits and precision (updated for 2003)

Asian Journal of Chemistry; Vol. 23, No. 7 (2011),

Determination of the concentration of caffeine, theobromine, and gallic acid in commercial tea samples

Determination of Metals in Wort and Beer Samples using the Agilent 5110 ICP-OES

CONCENTRATION LEVELS OF METALS IN COMMERCIALLY AVAILABLE ETHIOPIAN ROASTED COFFEE POWDERS AND THEIR INFUSIONS

The Determination of Pesticides in Wine

RESOLUTION OIV-OENO ANALYSIS OF VOLATILE COMPOUNDS IN WINES BY GAS CHROMATOGRAPHY

Application Note: Analysis of Melamine in Milk (updated: 04/17/09) Product: DPX-CX (1 ml or 5 ml) Page 1 of 5 INTRODUCTION

RIDASCREEN Gliadin. Validation Report. R-Biopharm AG. Art.No. R7001

Elemental Analysis of Yixing Tea Pots by Laser Excited Atomic. Fluorescence of Desorbed Plumes (PLEAF) Bruno Y. Cai * and N.H. Cheung Dec.

! " # # $% 004/2009. SpeedExtractor E-916

*Level IV report narratives are more detailed than other levels.

Speciated Arsenic Analysis in Wine Using HPLC-ICP-QQQ

Harvest Series 2017: Wine Analysis. Jasha Karasek. Winemaking Specialist Enartis USA

Levels of Cd, Pb, and Ni in Different Types of Vinegars

PECTINASE Product Code: P129

Fast Analysis of Smoke Taint Compounds in Wine with an Agilent J&W DB-HeavyWax GC Column

Allergens in wine a specific detection of Casein, Egg and Lysozyme

The Impact of Vineyard Origin and Winery on the Elemental Profile of Red Wines

EXTRACTION OF SEDIMENTS FOR AROMATIC AND CHLORINATED HYDROCARBONS

Extraction of Acrylamide from Coffee Using ISOLUTE. SLE+ Prior to LC-MS/MS Analysis

General overview of the two stages of the QuEChERS technique. Stage 1: Sample extraction. Stage 2: Sample cleanup

Uses of profiling trace metals in wine with ICP- MS and Mass Profiler Professional (MPP) for the wine industry

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

Keywords Green and black tea. Infusions. Sample preparation. Multi-element analysis. Principal component analysis. Linear discriminant analysis

Application Note FP High Sensitivity Coumarin Analysis. Introduction. Keywords

Analytical Method for Coumaphos (Targeted to agricultural, animal and fishery products)

Acta Chim. Slov. 2003, 50, THE CONTENTS OF Cu, Mn, Zn, Cd, Cr AND Pb AT DIFFERENT STAGES OF THE WINEMAKING PROCESS

Supporting Information

Comparison of Heating Extraction Procedures for Al, Ca, Mg, and Mn in Tea Samples

Validation Report: Total Sulfite Assay Kit (cat. no. K-TSULPH)

Room 6G,Floor 6,Building A1,Zhujiang Square,Zhongxin Cheng,Longgang District,Shenzhen,China

Bromine Containing Fumigants Determined as Total Inorganic Bromide

Application Note No. 184/2015

An Economic And Simple Purification Procedure For The Large-Scale Production Of Ovotransferrin From Egg White

CONTENT VARIATION OF IRON AND COPPER IN WINE OBTAINED FROM WINE VINEYARDS RECAŞ

QUANTITATIVE ASSAY FOR OCHRATOXIN A IN COFFEE, COCOA, AND SPICES (96-well kit)

Determination of Lead in Saudi Arabian Imported Green Tea by ICP-MS

In the preparation of this Tanzania Standard assistance was derived from:

GAS-CHROMATOGRAPHIC ANALYSIS OF SOME VOLATILE CONGENERS IN DIFFERENT TYPES OF STRONG ALCOHOLIC FRUIT SPIRITS

An Overview of Official Methods of Analysis

EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

High Sensitivity Quantitation Method of Dicyandiamide and Melamine in Milk Powders by Liquid Chromatography Tandem Mass Spectrometry

EXTRACTION OF SEDIMENTS FOR BUTYLTINS

RESOLUTION OIV-OENO MONOGRAPH ON GLUTATHIONE

Determination of Caffeine in Coffee Products According to DIN 20481

Laboratoire national de métrologie et d essais

EXTRACTION PROCEDURE

CORRELATIONS BETWEEN CUTICLE WAX AND OIL IN AVOCADOS

TSKgel TECHNICAL INFORMATION SHEET No. 131

Winemaking and Tartrate Instability

ASSET EZ4-NCO Dry Sampler Extraction Procedure.

Validation Report: Free Sulfite Assay Kit (cat. no. K-FSULPH)

Rapid Analysis of Soft Drinks Using the ACQUITY UPLC H-Class System with the Waters Beverage Analysis Kit

Alcolyzer Plus Spirits

powder and cocoa butter and various spices in the range of 1-20 ppb (µg/kg).

Solid Phase Micro Extraction of Flavor Compounds in Beer

DETERMINATION OF CADMIUM IN SAUDI ARABIAN IMPORTED GREEN TEA SAMPLES BY ICP-MS

Application Note. Hydrolysis Unit E-416, Extraction Unit E-816 Soxhlet. Fat determination according to Weibull-Stoldt - Standard application

COMPARISON OF FOUR MERLOT CLONAL SELECTIONS FROM SKOPJE S VINEYARD REGION, R. MACEDONIA

3. Aspirin Analysis. Prelaboratory Assignment. 3.1 Introduction

Samples: Standard solutions of rutin, quercetina, rosmarinic acid, caffeic acid and gallic acid. Commercial teas: Green, Ceilan, Hornimans and Black.

Spectrophotometric Flow- Injection Method for Determination of Sorbic Acid in Wines

Chapter 14 Tex-619-J, Analysis of Water for Chloride and Sulfate Ions

STUDIES ON THE CHROMATIC CHARACTERISTICS OF RED WINES AND COLOR EVOLUTION DURING MATURATION

Chemistry 212 MOLAR MASS OF A VOLATILE LIQUID USING THE IDEAL GAS LAW

Definition of Honey and Honey Products

Determination of Alcohol Content of Wine by Distillation followed by Density Determination by Hydrometry

Higher Resolution Separation of Organic Acids and Common Inorganic Anions in Wine

Solid Phase Micro Extraction of Flavor Compounds in Beer

Alcohol Meter for Wine. Alcolyzer Wine

Determination of Pesticides in Coffee with QuEChERS Extraction and Silica Gel SPE Cleanup

Unit Test: Nature of Science

ICP Test Report Certification Packet

Vinmetrica s SC-50 MLF Analyzer: a Comparison of Methods for Measuring Malic Acid in Wines.

Analysis of Beta-Carotene and Total Carotenoids from Pacific Sea Plasma (Spectrophotometric Method)

Determination of natamycin in wines Résolution OIV-SCMA

Somchai Rice 1, Jacek A. Koziel 1, Anne Fennell 2 1

Determination of Ochratoxin A in Roasted Coffee According to DIN EN 14132

CHEM Experiment 4 Introduction to Separation Techniques I. Objectives

One class classification based authentication of peanut oils by fatty

1. Blender: Osterizer, 10-speed, or equivalent. 2. Separatory Funnel: Kilborn or equivalent (see figure 1) 2. HCl Solution: HCl/water (7:93 by volume)

DEVELOPMENT OF A RAPID METHOD FOR THE ASSESSMENT OF PHENOLIC MATURITY IN BURGUNDY PINOT NOIR

Oregon Wine Advisory Board Research Progress Report

EXTRACTION. Extraction is a very common laboratory procedure used when isolating or purifying a product.

Fast Analysis of Arsenic Species in Wines using LC-ICP-QQQ

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

HI Formol Number Mini Titrator for Wine and Fruit Juice Analysis

The Cruel Exploiter- Acacia confusa (Taiwan Acacia)

A COMPARATIVE STUDY OF THE CAFFEINE PROFILE OF MATURE TEA LEAVES AND PROCESSED TEA MARKETED IN SONITPUR DISTRICT OF ASSAM, INDIA.

The delicate art of wine making. Alfa Laval Foodec decanter centrifuges in the wine industry

SUPPLEMENTARY MATERIALS. Methylxanthine content in commonly consumed foods in Spain and determination of its intake during consumption

Royal Society of Chemistry Analytical Division East Anglia Region National Schools' Analyst Competition

ALPHA. Innovation with Integrity. FT-IR Wine & Must Analyzer FT-IR

INFLUENCE OF THIN JUICE ph MANAGEMENT ON THICK JUICE COLOR IN A FACTORY UTILIZING WEAK CATION THIN JUICE SOFTENING

Transcription:

ETAAS DETERMINATION OF SOME TRACE ELEMENTS IN WINE T. STAFILOV A, J. CVETKOVIĆ B, S. ARPADJAN C, I. KARADJOVA C a Institute of Chemistry, Faculty of Science, Sts. Cyril and Methodius University, POB 162, 1000 Skopje, Macedonia; email: trajcest@iunona.pmf.ukim.edu.mk b Institute of Agriculture, Bull. A. Makedonski bb., 1000 Skopje, Macedonia c Faculty of Chemistry, University of Sofia, 1126 Sofia, Bulgaria Abstracts Methods for trace elements determination in wine by electrothermal atomic absorption spectrometry (ETAAS) directly in untreated wines and in previously decomposed wines are given. The investigations on the atomization behaviour of the microelements in wine, on the reasons for strong wine matrix interferences, on the accuracy of the calibration procedures are presented. Appropriate procedures for preliminary separation and preconcentration of ultra trace elements from wine matrix are compared and discussed. Optimal analytical methods for direct ETAAS determination of As, Cd, Cr, Ni, Pb and extraction ETAAS determination of Se, Tl are developed and validated. Results for the content of trace elements in Macedonian and Bulgarian wines are presented. Key words: trace elements, wine, determination, electrothermal atomic absorption spectrometry 1. Introduction The analysis of trace metals in wines is of great importance for quality control of wine, authenticity control of wine, wine metals toxicity, bioavailability etc. These elements might be essential or toxic in the human body, they can influence the wine making process or they can change the wine taste and quality. The presence of trace elements in wine is the consequence of the atmospheric deposition of airborne particulate matter on grapes and/or of the intake of these elements by the grapevine from ground water and soil, pesticides and fertilizers [1]. Trace elements in wine samples can be measured by different techniques: stripping voltammetry [29], instrumental neutron activation analysis [10], ICPAES, ICPMS [1116] or UVVis spectrophotometry [1720]. However official methods for the determination of heavy metals in wine produced by the Office International de la Vigne et du Vin and the American Society of Enologists are essentially based on atomic absorption spectrometry (AAS). Some elements with relatively high concentration in wine could be analyzed by flame atomic absorption spectrometry (FAAS) [2137]. But for many elements present in wine with low concentration, hydride generation techniques with FAAS (HGAAS) [3844] or mostly electrothermal atomic absorption spectrometry (ETAAS) [4577] is used. For some elements direct determination (or using simple dilution) of trace elements in wine by ETAAS was applied [50,53,5864,73]. However, very often some additional operations in the preparation of wine samples are necessary because of matrix interferences [5153,59,64,66,72,73]. In order to avoid these interferences and because of the very low concentration of trace elements in the samples investigated, it is necessary to use matrix modifier [45,48,55,56,6569] or to separate and concentrate these elements [45,46,77]. To reach this goal, different methods could be employed: extraction, ion exchange, precipitation, etc. In this work, the possibilities of electrothermal atomic absorption spectrometry for the determination of As, Cd, Cr, Ni, Pb, Se and Tl in different wines (white, red and rose) are evaluated. Optimal instrumental parameters (temperature programs, modifiers, atomizers) for direct ETAAS determination of As, Cd, Cr, Ni, Pb and extraction ETAAS determination of Se and Tl are presented. Suitable calibration procedures are proposed. The accuracy, precision 90

and selectivity of the developed methods were evaluated. Comparative analysis by using different analytical procedures and atomic absorption instruments were used as validation procedure. Results for the content of trace elements in Macedonian and Bulgarian wines are presented. 2. Experimental 2.1. Instrumentation The electrothermal atomic absorption measurements were carried out on a Perkin Elmer Zeeman 3030 spectrometer coupled with an HGA600 atomizer and on a Varian SpectraAA 880 spectrometer coupled with GTA100. The light sources were hollow cathode lamps for As, Cd, Cr, Pb, Ni, Se and Tl (Varian) and hollow cathode lamps for Cr, Ni, Pb and Tl (PerkinElmer) and elecrodeless lamps for Cd and Se (PerkinElmer), the spectral bandpass and the wavelengths used were as recommended by PerkinElmer and Varian. Uncoated graphite tubes and uncoated graphite tubes with pyrolytic platforms (PerkinElmer) and with centre fixed platforms (Varian) were used as atomizers. Solutions (10 µl) were introduced manually into the graphite furnace by using Eppendorff pipettes for the PerkinElmer instrument and by using autosampler for SpectraAA 880. The graphite furnace operating parameters are presented in Table 1. Only peak areas were used for quantification. Table 1. Optimal experimental conditions for ETAAS determination of As, Cd, Cr, Ni, Pb, Se and Tl in wine Element As white wine red wine Pretreatment temperature ( o C) Atomization temperature ( o C) Atomizer Modifier Sample Calibration D 2 Zeeman D 2 Zeeman pretreatment 1400 1400 2600 2600 Platform Platform 5 µg Pd 5 µg Pd direct digestion aqueous st.sol. standard add. Cd 300 400 1600 1000 Platform no modifier direct standard add. Cr 1300 1500 2600 2500 Wall no modifier direct aqueous st.sol Ni 800 1300 2400 2300 Wall no modifier direct aqueous st.sol Pb 400 700 2100 1500 Platform no modifier direct standard add. Se 1400 2400 Platform Ag***, digestion, standard add. Ni+Mg(NO) 3 extraction Tl 350 900 2400 1500 Platform Ag,* tartaric acid digestion, extraction standard add. * Varian SpectrAA 880; ** PerkinElmer Zeeman 3030; *** modifier for measurements by PerkinElmer 2.2. Reagents All reagents and solvents were of analyticalreagent grade (E. Merck). Aqueous standard solutions were prepared by appropriate dilution of 1000 µg ml 1 stock solutions, which were made from Titrisol concentrates (Merck, Darmstadt, Germany). The matrix modifiers used were 10 % aqueous solution of tartaric acid, 10 g l 1 Ag as silver nitrate, Ni+Mg(NO 3 ) 2 and 0.3 g l 1 Pd as tetrachloroammonium palladate(ii) (Fluka AG). Ammonium pyrrolidin1yl dithioformate [ammonium pyrrolidinedithiocarbamate (APDC)] as 2% aqueous solution was used. The organic solvent isobutylmethyl ketone (IBMK) was applied without additional purification. Doubly distilled water was used throughout. 3. Results and Discussion 91

3.1. Direct ETAAS determination of As, Cd, Pb, Cr and Ni in wine As a first step the possibilities for the direct ETAAS determination of the studied elements is investigated. Analytical methods for the determination of As, Cd, Pb, Cr and Ni could be developed only after careful optimization of the instrumental parameters (temperature programs, modifiers, calibration procedures) since natural content of these elements in wines is above their characteristic masses. In order to obtain optimal temperature programs pretreatment and atomization curves (wall and platform atomization) were constructed for As, Cd, Pb, Cr and Ni in the presence of different wine samples filtered and untreated wines, wines obtained after different treatment procedures and bottled wines obtained from the market. The optimal instrumental parameters obtained for both studied instruments are summarized in Table 1. Wine matrix thermally stabilizes Cd and Pb and no modifier is proposed for their determination. The thermal behavior of As was studied for spiked wines (50 ng/ml As as As(III)) only with SpectraAA 880. The most effective modifier ensuring highest lossfree pretreatment temperature in presence of wine is Pd (5 µg Pd). The wine matrix interference on the atomization of As, Cd, Pb, Cr and Ni was evaluated through the ratio of the slopes of the calibration curves obtained in presence of wine (b m ) and for aqueous standard solutions (b o ). For Cr and Ni this ratio is 0.98±0.03. The strongest matrix interference was observed for As determination in red wines, the slope ratio was lower than 0.5 under optimal instrumental parameters for the both instruments. For white wine this ratio for As is 0.92±0.05, for Cd and Pb 0.93±0.06. As can be expected platform atomization removes in some extend matrix interference in the analysis of As, Cd and Pb for all types of the investigated wines. But even in this case standard addition calibration should be used as calibration procedure. It should be mention that the behaviour of the standard addition is significantly different when it is previously prepared in the wine sample and when the addition is performed directly in the graphite furnace by using the possibilities of the autosampler. This behaviour is also significantly different for different types of wine and the depression is more pronounced for treated and bottled wines. Hence, the standard addition should be performed to each wine sample prior to injection into the graphite furnace. 3.2. ETAAS determination of As, Cd, Pb, Cr and Ni after wine digestion Wine matrix digestion followed by ETAAS measurement was performed in order to check the accuracy of the results obtained for Cd, Cr, Pb and Ni and the possibilities to decrease the interferences in the analysis of As. The experiments provided showed that a mixture of HNO 3 and H 2 O 2 ensures complete and relatively fast wine matrix decomposition. Digestion procedure was performed in open vessel (bakers covered with watch glasses), closed system under reflux and microwave digestion system (Milestone). The mixture of HNO 3 and H 2 O 2 plays a role of modifier for Cd and Pb and these elements are determined in absence of modifier. Palladium was used as modifier for As. The matrix interference was evaluated again through the ratio of the slopes of the calibration curves. As expected the digestion procedure equalizes the behavior of all types of the investigated wines. Still the matrix interferences in As determination in red wine are relatively high and the standard addition method should be used for calibration. For Cd, Cr Ni and Pb calibration against aqueous standards is possible. Parallel analysis performed for all studied elements in different wines showed very good agreement between direct ETAAS determination and ETAAS determination after digestion. The comparison of the results obtained with different digestion procedures confirmed the suggestion that there are no losses of volatile elements during the digestion in open vessel. 92

3.3. Extraction ETAAS determination of Se and Tl The natural content of Se and Tl in wines is bellow the detection limits of these elements achieved by ETAAS, therefore a preconcentration step should be used for their determination. Liquid/liquid extraction was chosen as simple, fast and suitable for routine application method for separation and preconcentration of Se and Tl from wine samples. It is well known that only Se(IV) forms extractable dithiocarbamate complex in presence of 1 mol l 1 HCl and both Tl(I) and Tl(III) form extractable dithiocarbamate complexes in wide ph range (ph 1 to ph 6). The degree of extraction of Se (IV) was studied directly from wine samples acidified to 1 mol l 1 HCl and from digested wine samples evaporated to dryness and acidified to 1 mol l 1 HCl. The degree of extraction directly from wine samples was lower than 30 %, but from digested wine samples it was in the range 95%99% which can be used for Se determination. For thallium determination relatively higher volumes of wine samples should be used for preconcentration and the degree of extraction of the dithiocarbamate complexes of Tl(I) and Tl(III) is in the range of 6568 % when extracted directly from wine and 7580 % from digested wines. Therefore the extraction system KIIBMK was investigated as another possibility for preconcentration of Tl. It was found that in presence of 0.5 mol l 1 KI the extraction of both Tl(I) and Tl(III) into IBMK is quantitative. Unfortunately the degree of extraction directly from wines is only around 70 %, therefore extraction from digested wine samples (degree of extraction 9597%) is proposed for routine determination of Tl in wines. It should be mention that thermal behaviour of organic species of Se and Tl extracted into IBMK is significantly different for both instruments studied. Silver is proposed as modifier for thermal stabilization of both Se extracted as dithiocarbamate complex and Tl extracted as iodide complex for Zeeman 3030. However for Varian the most appropriate modifier for Se stabilization in organic phase is Ni+Sr(NO 3 ) 2 and for Tl tartaric acid which ensures higher lossfree pretreatment temperature with highest sensitivity of measurement. The optimal instrumental parameters are depicted also in Table 1. 3.4. Wine analysis Different Macedonian and Bulgarian wines were analyzed for As, Cd, Cr, Ni, Pb, Se and Tl according to the analytical procedures developed. The results obtained as a mean range are presented in Table 2. Table 2. Concentration range of As, Cd, Cr, Ni, Se and Tl in Macedonian and Bulgarian wines (in µg ml 1 ) Element Macedonian wines Bulgarian wines Other wines White, n= 10 Red, n= 16 Red, n= 16 White, n= 9 320 As 610 8 12 812 59 320 Cd 0.16 0.14 0.10.4 0.070.3 0.250.7 Cr 931 1030 1030 1220 3060 Ni 918 1040 1040 815 3050 Pb 50250 50250 3090 2070 30100 Se 0.50.8 0.51.3 0.51.3 0.40.7 0.613 Tl 0.10.5 0.30.7 0.30.7 0.10.5 not recorded 93

References [1] B. W. Zoecklein, K. C. Fugelsang, B. H. Gump, F. S. Nury, Wine Analysis and Production, Chapman&Hall, New York, 1994. [2] J. Wang, S. Mannino, Analyst, 114 (1989) 643. [3] C. Marin, P. Ostapczuk, Fresenius J. Anal. Chem., 343 (1992) 881. [4] H. R. Eschnauer, P. Ostapczuk, WeinWiss., 47 (1992) 206. [5] M. T. Arcos, M. C. Ancin, J. C. Echeverria, A. Gonzales, J. J Garrido, J. Agric. Food Chem., 41 (1993) 2333. [6] P. Ostapczuk, Anal. Chim. Acta, 273 (1993) 35. [7] E. A. Zakharova, V. M. Pichugina, T. P. Tolmacheva, Zh. Anal. Khim., 51 (1996) 918. [8] S. Sanllorente, O. M. Cruz, M. J. Arcos, Analyst, 123 (1998) 513. [9] O. Domingez, S. Sanllorente, M. J. Arcos, Electroanalysis, 1 (1999) 1273. [10] S. May, J. Leroy, D. Piccot, G. Pinte, J. Radioanal. Chem., 72 (1982) 305. [11] H. Eschnauer, L. Jakob, H. Meierer, R. Neeb, Mikrochim. Acta, 3 (1989) 291. [12] J. Goossens, T. De Smaele, L. Moens, R. Dams, Fresenius J. Anal. Chem., 347 (1993) 119. [13] S. Augagneur, B. Medina, J. Bernard, Szpunar, R. Lobinski, J. Anal. At. Spectrom., 11 (1996) 713. [14] J. D. Greenough, H. P. Longerich, S. E. Jackson, Australian J. Grape Wine Res., 3 (1997) 75. [15] M. Y. PeresJordan, J. Soldevila, A. Salvador, A. Pastor, M. de la Guardia, J. Anal. At. Scpectrom., 13 (1998) 33. [16] S. Wangkarn, S. A. Pergantis, J. Anal. At. Spectrom., 14 (1999) 657. [17] G. Weber, G. Schwedt, Anal. Chim. Acta, 134 (1982) 81. [18] C. M. GarciaJahres, M. A. LageYusty, J. SimalLozano, Fresenius J. Anal. Chem., 338 (1990) 703. [19] K. Ohzeki, I. Nukatsuka, K. Ichimura, F. Kumagai, M. Kogawa, Microchem. J., 49 (1994) 256. [20] S. Cherubin, S. Buiatti, F. Battistutta, R. Zironi, WeinWiss., 49 (1994) 78. [21] F. Bermejo Martinez, C. Baluja Santos, R. A. Ravina Pereiro, Analusis, 3 (1975) 157. [22] J. Korkisch, A. Sorio, I. Steffan, Talanta, 23 (1976) 289. [23] U. Anders, G. Hailer, Fresenius Z. Anal. Chem., 278 (1976) 203. [24] D. R. Yan, E. Stumpp, G. Schwedt, Fresenius Z. Anal. Chem., 322 (1985) 474. [25] R. Farre, M. J. Lagarda, J. Micronutr. Anal., 2 (1986) 201. [26] R. Barbera, R. Farre, Ann. Falsif. Expert. Chim. Toxicol., 79 (1986) 209. [27] H. Chow, B. H. Gump, J. Assoc. Off. Anal. Chem., 70 (1987) 61. [28] C. BalujaSantos, A. GonzalesPortal, J. M. BouzasBouzas, Anal. Chim. Acta, 196 (1987) 283. [29] J. Hernandez Mendez, O. Jimenez de Blas, A. Del Valle Gonzalez, D. Alonso Gutierrez, Food Chem., 31 (1989) 205. [30] R. Ajlec, J. Stupar, Analyst, 114 (1989) 137. [31] G. Weber, Fresenius J. Anal. Chem., 340 (1991) 161. [32] A. J. McKinnon, R. W. Cattrall, G. R. Scollary, Am. J. Enol. Vitic., 43 (1992) 166. [33] M. Netzer, F. Bandion, Mitt. Klosterneuburg, 42 (1992) 250. [33] R. A. S. Lapa, J. L. F. C. Jose, J. L. M. Santos, Food Chem., 55 (1996) 397. [34] K. Liftinger, M. Netzer, F. Bandion, Mitt. Klosterneuburg, 46 (1996) 162. [35] F. A. Chmilenko, L. V. Baklanova, Zh. Anal. Khim., 52 (1997) 1093. [36] M. Olalla, M. C. Gonzales, C. Cabrera, M. C. Lopez, J. AOAC Int., 83 (2000) 189. [37] R. Stefani, Lab., 9 (2000) 5055. [38] D. D. Siemer, R. K. Vitek, P. Koteel, H. Prabhakaran, W. C. Houser, Anal. Lett., 10 (1977) 357. [39] J. Sanz, P. Basterra, J. Galban, J. R. Castillo, Mikrochim. Acta, 1 (1989) 271. [40] J. Sanz, P. Basterra, J. Galban, J. R. Castillo, Microchem. J., 40 (1989) 115. [41] J. Cacho, V. Ferreira, C. Nerin, Analyst, 117 (1992) 31. [42] C. Cabrera, Y. Madrid, C. Camara, J. Anal. At. Spectrom., 9 (1994) 1423. [43] C. M. Mena, C. Cabrera, M. Luisa Lorenza, M. Carmen Lopez, J. Agric. Food Chem., 45 (1987) 1812 1815. [44] M. Segura, Y. Madrid, C. Camara, J. Anal. At. Spectrom., 14 (1999) 131135. [45] W. Ooghe, H. Kastelijn, Ann. Falsif. Expert. Chim., 69 (1976) 35167. [46] P. Petrino, M. Cas, J. Estienne, Ann. Falsif. Expert. Chim., 69 (1976) 87. [47] G. Severin, E. Schumacher, F. Umland, Fresenius Z. Anal. Chem., 311 (1982) 2058. [48] G. Severin, E. Schumacher, F. Umland, Fresenius Z. Anal. Chem., 311 (1982) 2058. [49] G. P. Marletta, L. G. Favretto, J. Sci. Food Agric., 37 (1986) 10916. [50] B. Casseta, Boll. Chim. Ig. Parte Sci., 35 (1986) 28794. [51] J. Cacho, J. Enrique Castella, At. Spectrosc., 10 (1989) 85. [52] Y. Wang, C. Lu, Z. Xiao, G. Wang, S. S. Kua, E. J. Rigsby, J. Agric. Food Chem., 39 (1991) 724. 94

[53] P. L. Teissedre, M. T. Cabanis, J. C. Cabanis, Analusis, 21 (1993) 249. [54] A. A. Almeida, M. Lourdes Bastos, M. Isabel Cardoso, M. A. Ferreira, J. L. F. C. Lima, M. Elisa Soares, J. Anal. At. Spectrom., 7 (1992) 1281. [55] R. M. Mattews, P. J. Parsons, At. Spectrosc., 14 (1993) 41. [56] M. Froning, C. Mohl, T. Ostapczuk, Fresenius J. Anal. Chem., 345 (1993) 233. [57] W. R. Mindak, J. AOAC Int., 77 (1994) 1023. [58] M. Larroque, J. C. Cabanis, J. AOAC Int., 77 (1994) 463. [59] S. N. Bruno, R. C. Campos, A. J. Curtius, J. Anal. At. Spectrom., 9 (1994) 341. [60] A. A. Almeida, I. M. Cardoso, At. Spectrosc., 15 (1994) 73. [61] A. Escobal, C. Iriondo, C. Laborra, E. Ulibarrena, At. Spectrosc., 16 (1995) 162. [62] L. Jorhem, B. Sundstroem, At. Spectrosc., 16 (1995) 226. [63] M. E. Soares, M. L. Bastos, M. A. Ferreira, At. Spectrosc., 16 (1995) 256. [64] B. T. Kildahl, W. Lund, Fresenius J. Anal. Chem., 354 (1996) 93. [65] C. Mena, C. Cabrera, M. L. Lorenzo, M. C. Lopez, Sci. Total Environ., 181 (1996) 201. [66] M. LopezArtiguez, A. M. Camean, M. Repetto, J. AOAC Int., 79 (1996) 1191. [67] V. I. Slaveykova, D. L. Tsalev, R. Maskill, Cinecia (Maracaibo), 4 (1996) 165172. [68] P. A. Brereton, C. M. Sargent, H. M. Crews, R. Wood, J. AOAC Int., 80 (1997) 1287. [69] J. Jaganathan, A. L. Reisig, S. M. Dugar, Microchem. J., 56 (1997) 221. [70] A.M., Wifladt, G. Wibetoe, W. Lund, Fresenius' J. Anal. Chem., 357 (1997) 92. [71] A. A. Almeida, M. I. Cardoso, J. L. F. C. Lima, J. Anal. At. Spectrom., 12 (1997) 837. [72] C. CabreraVique, P.L. Teissedre, M.T. Cabanis, J.C. Cabanis, J. AOAC Int., 80 (1997) 57. [73] C. CabreraVique, P.L. Teissedre, M.T. Cabanis, J.C. Cabanis, J. Agric. Food Chem., 45 (1997) 1808. [74] M. Segura, J. Grgic, Z. Grgic, B. Segura, Dtsch. Lebensm.Rundsch., 94 (1998) 336. [75] E. Lendinez, M. C. Lopez, C. Cabrera, M. L. Lorenzo, J. AOAC Int., 81 (1998) 1043. [76] C. CabreraVigne, P. L. Teissedre, M. T. Cabanis, J. C. Cabanis, Am. J. Enol. Vitic., 51 (2000) 103. [77] J. Cvetkovic, T. Stafilov, D. Mihajlovic, Fresenius' J. Anal. Chem., (2001), in press. 95