Validation Report: Free Sulfite Assay Kit (cat. no. K-FSULPH)

Similar documents
Validation Report: Total Sulfite Assay Kit (cat. no. K-TSULPH)

Measuring tannins in grapes and red wine using the MCP (methyl cellulose precipitable tannin assay

RIDASCREEN Gliadin. Validation Report. R-Biopharm AG. Art.No. R7001

Rapid Analysis of Soft Drinks Using the ACQUITY UPLC H-Class System with the Waters Beverage Analysis Kit

An Overview of Official Methods of Analysis

Allergens in wine a specific detection of Casein, Egg and Lysozyme

Application Note: Analysis of Melamine in Milk (updated: 04/17/09) Product: DPX-CX (1 ml or 5 ml) Page 1 of 5 INTRODUCTION

Determination of Caffeine in Coffee Products According to DIN 20481

Correlation of the free amino nitrogen and nitrogen by O-phthaldialdehyde methods in the assay of beer

Vinmetrica s SC-50 MLF Analyzer: a Comparison of Methods for Measuring Malic Acid in Wines.

Determination of Methylcafestol in Roasted Coffee Products According to DIN 10779

Laboratory Performance Assessment. Report. Analysis of Pesticides and Anthraquinone. in Black Tea

HOW MUCH DYE IS IN DRINK?

Royal Society of Chemistry Analytical Division East Anglia Region National Schools' Analyst Competition

Extraction of Acrylamide from Coffee Using ISOLUTE. SLE+ Prior to LC-MS/MS Analysis

Identification of Adulteration or origins of whisky and alcohol with the Electronic Nose

AgraQuant F.A.S.T. Egg. Test Kits available: AgraQuant. AgraQuant F.A.S.T. Cashew. AgraQuant F.A.S.T. Peanut

The AgraQuant Plus Allergen. Test Kits available: AgraQuant. AgraQuant Walnut. AgraQuant Plus Macadamia nut. AgraQuant Allergen Test Kits available:

EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

PRIMARY AMINO NITROGEN (PAN) ASSAY PROCEDURE

Extraction of Multiple Mycotoxins From Animal Feed Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

DEVELOPMENT OF A RAPID METHOD FOR THE ASSESSMENT OF PHENOLIC MATURITY IN BURGUNDY PINOT NOIR

AllerFlow Gluten Validation & Comparison

Bromine Containing Fumigants Determined as Total Inorganic Bromide

Determination of Ochratoxin A in Roasted Coffee According to DIN EN 14132

Analysis of trace elements and major components in wine with the Thermo Scientific icap 7400 ICP-OES

RESOLUTION OIV-OENO MONOGRAPH ON GLUTATHIONE

Application Note FP High Sensitivity Coumarin Analysis. Introduction. Keywords

Analytical Method for Coumaphos (Targeted to agricultural, animal and fishery products)

Application & Method. doughlab. Torque. 10 min. Time. Dough Rheometer with Variable Temperature & Mixing Energy. Standard Method: AACCI

Methanol (Resolution Oeno 377/2009, Revised by OIV-OENO 480/2014)

HI Formol Number Mini Titrator for Wine and Fruit Juice Analysis

Analysis Report Wine-ProfilingTM

Measuring Sulfur Dioxide: A Perennial Issue. Tom Collins Fosters Wine Estates Americas

Specify the requirements to be met by agricultural Europe Soya soya bean collectors and Europe Soya primary collectors.

806 West Beacon Rd. Client Project #: PCSB-LEAD First Draw Lakeland,FL Date Sampled: Oct 3, 2018 Oct 9, 2018; Invoice:

Winemaking and Sulfur Dioxide

OenoFoss. Instant quality control throughout the winemaking process. Dedicated Analytical Solutions

Allergen analysis of food and surfaces with sensitive test kits

Determination of natamycin in wines Résolution OIV-SCMA

Fact Sheet ANALYSIS. Measuring total anthocyanins (colour) in red grape berries. Scope. Equipment and apparatus

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

DOMESTIC MARKET MATURITY TESTING

Appendix B: Monitoring methods, accuracy, detection limits and precision (updated for 2003)

AAFCO Meeting August, Moisture Best Practices By Dr. Jane Caldwell Assistant Technical Director Midwest Laboratories

MALT & BACTERIAL BETA-GLUCANASE & CELLULASE

Assessment of the CDR BeerLab Touch Analyser. March Report for: QuadraChem Laboratories Ltd. Campden BRI Group contracting company:

Buying Filberts On a Sample Basis

Alcolyzer Plus Spirits

Use of a CEP. CEP: What does it mean? Pascale Poukens-Renwart. Certification of Substances Department, EDQM

Beer bitterness and testing

The Determination of Pesticides in Wine

R-Biopharm Inc. Allergens. Product overview. R-Biopharm for the safety of your analysis.

Food Allergen and Adulteration Test Kits

ISO INTERNATIONAL STANDARD. Oilseed residues Determination of oil content Part 2: Rapid extraction method

Application Note No. 193/2015

myclobutanil 987 MYCLOBUTANIL (181)

Application Note CL0311. Introduction

ANALYTICAL SUMMARY REPORT

Elemental Analysis of Yixing Tea Pots by Laser Excited Atomic. Fluorescence of Desorbed Plumes (PLEAF) Bruno Y. Cai * and N.H. Cheung Dec.

Wine analysis to check quality and authenticity by fully-automated 1

Volume NaOH ph ph/ Vol (ml)

Extraction of Phenolic Acids from Plant Tissue Using Accelerated Solvent Extraction (ASE )

Almond ß-Lactoglobulin (BLG) Casein Egg Gliadin (Gluten) Hazelnut Lupine Mustard Peanut Sesame Crustacea Soy Total Milk (Casein & Whey) Walnut

Thought Starter. European Conference on MRL-Setting for Biocides

QUANTITATIVE ASSAY FOR OCHRATOXIN A IN COFFEE, COCOA, AND SPICES (96-well kit)

LABORATORY PRACTICES IN WINE ANALYSIS. Dpto. Nutrición y Bromatología II. Facultad de Farmacia. UCM

ISO 712 INTERNATIONAL STANDARD. Cereals and cereal products Determination of moisture content Reference method

SUPPLEMENTARY MATERIALS. Methylxanthine content in commonly consumed foods in Spain and determination of its intake during consumption

Setting up your fermentation

Towards a numerical phenotyping for: Phenology Berry enological traits

RESOLUTION OIV-OENO ANALYSIS OF VOLATILE COMPOUNDS IN WINES BY GAS CHROMATOGRAPHY

Recent Developments in Rheological Instruments

REPORT OF THE WORKING GROUP ON TEA BREW BY T C CHAUDHURI N MURALEEDHARAN ANOOP KUMAR BAROOAH

The Brabender GlutoPeak A new type of dough rheology

Introduction to Measurement and Error Analysis: Measuring the Density of a Solution

Meeting Your GLUTEN Testing Needs

powder and cocoa butter and various spices in the range of 1-20 ppb (µg/kg).

THE WINEMAKER S TOOL KIT UCD V&E: Recognizing Non-Microbial Taints; May 18, 2017

Mastering Measurements

Use of Accelerated Solvent Extraction (ASE) with Centrifugal Evaporation to Automate Fat Determination in Chocolate

PROFICIENCY TESTS NO 19 AND EURL-Campylobacter National Veterinary Institute

The first and only fully-automated, random access, multiplex solution for Celiac IgA and Celiac IgG autoantibody testing.

Quantification of Asbestos in Soils

AWRI Refrigeration Demand Calculator

Alcoholic Fermentation in Yeast A Bioengineering Design Challenge 1

ab Anti-Deamidated Gliadin Peptide (DGP) IgG ELISA Kit

The Wild Bean Population: Estimating Population Size Using the Mark and Recapture Method

Evaluation of winemaking treatments in Australian Cabernet Sauvignon. Vintage trial 2018

Application Note No. 184/2015

Thermal Properties and Temperature

Detection and quantitation of various food allergens by LC-MS/MS

J. M. C. Dang 1 and M. L. Bason 1,2

The first and only fully-automated, random access, multiplex solution for Celiac IgA and Celiac IgG autoantibody testing.

Alcohol Meter for Wine. Alcolyzer Wine

The European Commission s science and knowledge service

LEVEL 4 AWARD IN FOOD SAFETY MANAGEMENT in MANUFACTURING

NEAR INFRARED SPECTROSCOPY (NIR) -SPECTROSCOPY, COLOUR MEASUREMENT AND SINGLE KERNEL CHARACTERIZATION IN RYE BREEDING

Innovations for a better world. Ingredient Handling For bakeries and other food processing facilities

Transcription:

Validation Report: Free Sulfite Assay Kit (cat. no. K-FSULPH) 1. Scope Megazyme s Free Sulfite Assay Kit (K-FSULPH) is a reliable and accurate method used for the rapid measurement and analysis of total sulfite (sulfite) in wine, beverages, foodstuffs and other materials. This method measures Free sulfite in mg/l and is used widely in the wine industry. This method has been validated for red and white wines at the Bundesamt für Weinbau, Austria. 2. Planning The purpose of this report is to verify and validate the current method as detailed by Free Sulfite Assay Kit (K-FSULPH). 3. Performance characteristics The selectivity, working range, limit of detection, limit of quantification, trueness (bias) and precision of this kit is detailed in this report. 3.1. Selectivity This assay is specific for free sulfite. Thiol reactive compounds (e.g. aldehydes, maleimide compounds) can interfere with the FSO2 assay. Aldehyde levels in wine do not cause interference of the FSO2 assay. 3.2. Working Range Assay follows the Free Sulfite Assay Kit (K-FSULPH) standard procedure. 0.05 ml of sulfite standard was used as sample, with a range of concentrations (5-150 mg/l sulfite) which corresponds to 0.25-7.5 µg of sulfite per cuvette. Absorbance A2 was read after 3 min, and A3, after a subsequent 3 min, at 575 nm and at 25 o C as recommended in the procedure. 1

Absorbance @ 575 nm Megazyme, Free Sulfite Concentration [μg/assay] 0 0.000 A 575nm 1.25 0.251 2.5 0.496 5.0 0.972 6.25 1.232 7.5 1.507 Free Sulfite Standard Curve 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0-0.2 R² = 0.9995 0 1 2 3 4 5 6 7 8 Free Sulfite (μg per assay) 2

3.3. LOD and LOQ The instrument limit of detection, as per kit booklet, is ~ 2 mg/l of sulfite, which is derived from an absorbance difference of 0.020 with the maximum sample volume of 0.05 ml. The calculated limit of detection (LOD) and the calculated limit of quantification (LOQ) for this report purpose is based on the analysis of samples that have been taken through the whole Free Sulfite Assay Kit (K-FSULPH) measurement procedure. The LOD is the lowest concentration of the analyte that can be detected by the method. LOD is calculated as 3 x s 0; where s 0 is the standard deviation of a number of samples A1 reading. The LOQ is the lowest level at which the kit s performance is acceptably repeatable. LOQ is calculated as kq x s 0; where s 0 is the standard deviation of a number of samples A1 reading. The IUPAC default value for kq is 10. For Free Sulfite Assay Kit (K-FSULPH) LOD For 0.05 ml of sample (maximum volume) Free Sulfite = 1.090 mg/l LOQ For 0.05 ml of sample (maximum volume) Free Sulfite = 2.280 mg/l * Note: The above detection limits are for samples as used in the assay, after sample preparations if required (e.g. deproteinisation). The dilution used in pre-treatment must be accounted for while establishing the detection limits for specific samples. 3

3.4. Trueness (Bias) Comparison of the mean of the results (x) achieved with the Free Sulfite Assay Kit (K- FSULPH) method with a suitable reference value (x ref). For this report, Relative Bias is calculated in per cent as: b(%) = x xref / xref x 100. The reference material for this purpose is sodium bisulphite supplied with the Free Sulfite Assay Kit (K-FSULPH) as FSO2 at 150 mg/l. Relative Bias b(%) n Ref Material Mean b(%) Free Sulfite 10 150 156.02 4.01 3.5. Precision This report details the reproducibility of the Free Sulfite Assay Kit (K-FSULPH), it is a measure of the variability in results, on different days and by different analysts, over an extended period of time. For the purpose of this report different lot numbers of the kit standard are used as the reference material. Reproducibility n Ref Material Mean Standard Deviation %CV Free Sulfite 10 150 156.02 7.4025 4.74 4

Repeatability of this kit can be assessed using wine samples. This is a measure of the variability in results by a single analyst, using real samples, using the same equipment and over a short period of time. The use of wine samples shows one of the many applications of this kit. Repeatability n Mean Standard Deviation %CV White Wine 12 22.43 1.07 4.76 Red Wine 12 43.16 1.29 2.98 4. Conclusion The method outlined in this document is a robust, quick and easy method for the measurement of free sulfite in various matrices. It has been used for many years and is fully automatable for high throughput analysis of samples. Data presented in this report verifies and validates that this method is fit for the purpose intended, which is summarised below. Validation Summary Free Sulfite Working range (µg in cuvette) 0.25-7.5 LOD 1.090 LOQ 2.280 Relative Bias b (%) 4.01 Reproducibility (%CV using sodium bisulfite) 4.74 Repeatability (%CV using white wine) 4.76 Repeatability (%CV using red wine) 2.98 5